Optimal threshold of postintervention minimum stent area to predict in-stent restenosis in small coronary arteries: An optical coherence tomography analysis Spontaneous Coronary Artery Dissection: Pathophysiological Insights From Optical Coherence Tomography Vascular response and healing profile of everolimus-eluting bioresorbable vascular scaffolds for percutaneous treatment of chronic total coronary occlusions: A one-year optical coherence tomography analysis from the GHOST-CTO registry Clinical Predictors for Lack of Favorable Vascular Response to Statin Therapy in Patients With Coronary Artery Disease: A Serial Optical Coherence Tomography Study Optical Coherence Tomography Guided Percutaneous Coronary Intervention With Nobori Stent Implantation in Patients With Non-ST-Segment-Elevation Myocardial Infarction (OCTACS) Trial: Difference in Strut Coverage and Dynamic Malapposition Patterns at 6 Months Optical Coherence Tomography to Optimize Results of Percutaneous Coronary Intervention in Patients with Non-ST-Elevation Acute Coronary Syndrome: Results of the Multicenter, Randomized DOCTORS Study (Does Optical Coherence Tomography Optimize Results of Stenting) Angiography alone versus angiography plus optical coherence tomography to guide decision-making during percutaneous coronary intervention: the Centro per la Lotta contro l'Infarto-Optimisation of Percutaneous Coronary Intervention (CLI-OPCI) study Fate of post-procedural malapposition of everolimus-eluting polymeric bioresorbable scaffold and everolimus-eluting cobalt chromium metallic stent in human coronary arteries: sequential assessment with optical coherence tomography in ABSORB Japan trial

Clinical Trial2013 Oct;6(10):1095-1104.

JOURNAL:JACC Cardiovasc Imaging. Article Link

OCT compared with IVUS in a coronary lesion assessment: the OPUS-CLASS study

Kubo T, Akasaka T, Shite J et al. Keywords: FD; IVUS; MLA; MLD; OCT; PCI; QCA; TD; coronary angiography; frequency domain; intravascular ultrasound; minimum lumen area; minimum lumen diameter; optical coherence tomography; percutaneous coronary intervention; quantitative coronary angiography; time-domain

ABSTRACT


OBJECTIVESThe aim of this study was to investigate the reliability of frequency domain optical coherence tomography (FD-OCT) for coronary measurements compared with quantitative coronary angiography (QCA) and intravascular ultrasound (IVUS).


BACKGROUND - Accurate luminal measurement is expected in FD-OCT because this technology offers high resolution and excellent contrast between lumen and vessel wall.

METHODS - In 5 medical centers, 100 patients with coronary artery disease were prospectively studied by using angiography, FD-OCT, and IVUS. In addition, 5 phantom models of known lumen dimensions (lumen diameter 3.08 mm; lumen area 7.45 mm(2)) were examined using FD-OCT and IVUS. Quantitative image analyses of the coronary arteries and phantom models were performed by an independent core laboratory.

RESULTS - In the clinical study, the mean minimum lumen diameter measured by QCA was significantly smaller than that measured by FD-OCT (1.81 ± 0.72 mm vs. 1.91 ± 0.69 mm; p < 0.001) and the minimum lumen diameter measured by IVUS was significantly greater than that measured by FD-OCT (2.09 ± 0.60 mm vs. 1.91 ± 0.69 mm; p < 0.001). The minimum lumen area measured by IVUS was significantly greater than that by FD-OCT (3.68 ± 2.06 mm(2) vs. 3.27 ± 2.22 mm(2); p < 0.001), although a significant correlation was observed between the 2 imaging techniques (r = 0.95, p < 0.001; mean difference 0.41 mm(2)). Both FD-OCT and IVUS exhibited good interobserver reproducibility, but the root-mean-squared deviation between measurements was approximately twice as high for the IVUS measurements compared with the FD-OCT measurements (0.32 mm(2) vs. 0.16 mm(2)). In a phantom model, the mean lumen area according to FD-OCT was equal to the actual lumen area of the phantom model, with low SD; IVUS overestimated the lumen area and was less reproducible than FD-OCT (8.03 ± 0.58 mm(2) vs. 7.45 ± 0.17 mm(2); p < 0.001).

CONCLUSIONS - The results of this prospective multicenter study demonstrate that FD-OCT provides accurate and reproducible quantitative measurements of coronary dimensions in the clinical setting.