The European bifurcation club Left Main Coronary Stent study: a randomized comparison of stepwise provisional vs. systematic dual stenting strategies (EBC MAIN) ‘Small bifurcation?’ CT myocardial mass volume measurements change therapeutic strategy in coronary artery disease T and small protrusion (TAP) vs double kissing crush technique: Insights from in-vitro models Effect of Side Branch Predilation in Coronary Bifurcation Stenting With the Provisional Approach - Results From the COBIS (Coronary Bifurcation Stenting) II Registry Incidence of Adverse Events at 3 Months Versus at 12 Months After Dual Antiplatelet Therapy Cessation in Patients Treated With Thin Stents With Unprotected Left Main or Coronary Bifurcations Double-Kissing Culotte Technique for Coronary Bifurcation Stenting - Technical evaluation and comparison with conventional double stenting techniques In vitro flow and optical coherence tomography comparison of two bailout techniques after failed provisional stenting for bifurcation percutaneous coronary interventions Effect of low-density lipoprotein cholesterol on the geometry of coronary bifurcation lesions and clinical outcomes of coronary interventions in the J-REVERSE registry

Original Research2018 Aug;31(4):421-429.

JOURNAL:J Interv Cardiol. Article Link

Impact of stent deformity induced by the kissing balloon technique for bifurcating lesions on in-stent restenosis after coronary intervention

Gamou T, Sataka K,Yamagishi M et al. Keywords: coronary bifurcation; intravascular ultrasound; kissing balloon technique

ABSRACT



OBJECTIVES - To investigate the impact of stent deformity induced by final kissing balloon technique (KBT) for coronary bifurcation lesions on in-stent restenosis (ISR).

 

BACKGROUND - In experimental models, the detrimental effects of KBT have been clearly demonstrated, but few data exists regarding the impact of proximal stent deformity induced by KBT on clinical outcomes.

 

METHODS - We examined 370 coronary lesions where intravascular ultrasound (IVUS)-guided second-generation drug-eluting stent (DES) implantation for coronary bifurcation lesions was performed. Based on IVUS analysis, the stent symmetry index (minimum/maximum stent diameter) and stent overstretch index (the mean of stent diameter/the mean of reference diameter) were calculated in the proximal main vessel.

 

RESULTS - The stent symmetry index was significantly lower (0.75 ± 0.07 vs 0.88 ± 0.06, P < 0.0001) and the stent overstretch index was significantly higher (1.04 ± 0.08 vs 1.01 ± 0.06, P = 0.0007) in lesions with KBT (n = 174) compared to those without KBT (n = 196). The number of two-stent technique in lesions with KBT was 31 (18%). In multivariate analysis, the degree of stent deformity indices was not associated with ISR in lesions with KBT; however, two-stent technique use was the only independent predictor of ISR at 8 months (hazard ratio: 3.96, 95% confidence interval: 1.25-12.5, P = 0.01).

 

CONCLUSIONS - Second-generation DES deformity induced by KBT was not associated with mid-term ISR.

 

 