Dabigatran dual therapy with ticagrelor or clopidogrel after percutaneous coronary intervention in atrial fibrillation patients with or without acute coronary syndrome: a subgroup analysis from the RE-DUAL PCI trial Anticoagulation in Concomitant Chronic Kidney Disease and Atrial Fibrillation: JACC Review Topic of the Week Comparison of the Efficacy and Safety Outcomes of Edoxaban in 8040 Women Versus 13 065 Men With Atrial Fibrillation in the ENGAGE AF-TIMI 48 Trial Extracellular Vesicles From Epicardial Fat Facilitate Atrial Fibrillation Percutaneous Left Atrial Appendage Closure for Stroke Prophylaxis in Patients With Atrial Fibrillation: 2.3-Year Follow-up of the PROTECT AF (Watchman Left Atrial Appendage System for Embolic Protection in Patients With Atrial Fibrillation) Trial Subclinical and Device-Detected Atrial Fibrillation: Pondering the Knowledge Gap: A Scientific Statement From the American Heart Association Early Rhythm-Control Therapy in Patients with Atrial Fibrillation Edoxaban-based versus vitamin K antagonist-based antithrombotic regimen after successful coronary stenting in patients with atrial fibrillation (ENTRUST-AF PCI): a randomised, open-label, phase 3b trial Left atrial appendage occlusion in atrial fibrillation patients with previous intracranial bleeding: A national multicenter study EHRA/EAPCI expert consensus statement on catheter-based left atrial appendage occlusion – an update
|<< 1 2 3 4 >>|

Clinical TrialVolume 12, Issue 24, December 2019

JOURNAL:JACC Cardiovasc Interv. Article Link

Left Ventricular Rapid Pacing Via the Valve Delivery Guidewire in Transcatheter Aortic Valve Replacement

Benjamin Faurie, Géraud Souteyrand, the EASY TAVI Investigators. Keywords: left ventricular pacing; left ventricular stimulation; transcatheter aortic valve implantation; transcatheter aortic valve replacement

ABSTRACT

OBJECTIVES -  This study investigated whether left ventricular (LV) stimulation via a guidewire-reduced procedure duration while maintaining efficacy and safety compared with standard right ventricular (RV) stimulation.

 

BACKGROUND -  Rapid ventricular pacing is necessary to ensure cardiac standstill during transcatheter aortic valve replacement (TAVR).

 

METHODS -  This is a prospective, multicenter, single-blinded, superiority, randomized controlled trial. Patients undergoing transfemoral TAVR with a SAPIEN valve (Edwards Lifesciences, Irvine, California) were allocated to LV or RV stimulation. The primary endpoint was procedure duration. Secondary endpoints included efficacy, safety, and cost at 30 days.

 

RESULTS -  Between May 2017 and May 2018, 307 patients were randomized, but 4 were excluded because they did not receive the intended treatment: 303 patients were analyzed in the LV (n = 151) or RV (n = 152) stimulation groups. Mean procedure duration was significantly shorter in the LV stimulation group (48.4 ± 16.9 min vs. 55.6 ± 26.9 min; p = 0.0013), with a difference of 0.12 (95% confidence interval: 0.20 to 0.05) in the log-transformed procedure duration (p = 0.0012). Effective stimulation was similar in the LV and RV stimulation groups: 124 (84.9%) versus 128 (87.1%) (p = 0.60). Safety of stimulation was also similar in the LV and RV stimulation groups: procedural success occurred in 151 (100%) versus 151 (99.3%) patients (p = 0.99); 30-day MACE-TAVR (major adverse cardiovascular event-transcatheter aortic valve replacement) occurred in 21 (13.9%) versus 26 (17.1%) patients (p = 0.44); fluoroscopy time (min) was lower in the LV stimulation group (13.48 ± 5.98 vs. 14.60 ± 5.59; p = 0.02), as was cost (18,807 ± 1,318 vs. 19,437 ± 2,318; p = 0.001).

 

CONCLUSIONS - Compared with RV stimulation, LV stimulation during TAVR was associated with significantly reduced procedure duration, fluoroscopy time, and cost, with similar efficacy and safety. (Direct Left Ventricular Rapid Pacing Via the Valve Delivery Guide-wire in TAVR [EASY TAVI]; NCT02781896)