Half-Dose Direct Oral Anticoagulation Versus Standard Antithrombotic Therapy After Left Atrial Appendage Occlusion Left Atrial Appendage Closure versus Non-Warfarin Oral Anticoagulation in Atrial Fibrillation: 4-Year Outcomes of PRAGUE-17 Patent Foramen Ovale Attributable Cryptogenic Embolism With Thrombophilia Has Higher Risk for Recurrence and Responds to Closure Procedural and Short-Term Results With the New Watchman FLX Left Atrial Appendage Occlusion Device Does pulsed field ablation regress over time? A quantitative temporal analysis of pulmonary vein isolation Transseptal puncture versus patent foramen ovale or atrial septal defect access for left atrial appendage closure 2015 ACC/HRS/SCAI Left Atrial Appendage Occlusion Device Societal Overview Expert Recommendations on Cardiac Computed Tomography for Planning Transcatheter Left Atrial Appendage Occlusion Percutaneous left atrial appendage occlusion: the Munich consensus document on definitions, endpoints, and data collection requirements for clinical studies Dabigatran dual therapy with ticagrelor or clopidogrel after percutaneous coronary intervention in atrial fibrillation patients with or without acute coronary syndrome: a subgroup analysis from the RE-DUAL PCI trial

Original Research2020 Jul 14;EIJ-D-20-00334.

JOURNAL:Eurointervention. Article Link

Feasibility and efficacy of the ultrashort side branch dedicated balloon in coronary bifurcation stenting

Y Murasato, M Nishihara, T Mori et al. Keywords: kissing balloon inflation vs. glider balloon; stent configuration; high-angled bifurcation; coronary bifurcation stenting

ABSTRACT

AIMS - We sought to investigate the feasibility and efficacy of the Glider balloon (GB), a side branch (SB) dedicated balloon 4-mm in length, in coronary bifurcation stenting.

METHODS AND RESULTS - In bifurcation bench models, stent configuration was examined with micro-focus computed tomography after crossover stenting followed by GB dilation or kissing balloon inflation (KBI). GB dilation maintained cross-sectional stent area without significant deformation and presented effective jailed strut removal in a high-angled bifurcation model.


We performed GB dilatation after main vessel (MV) stenting for 207 lesions in 194 patients, which included left main, true-bifurcation lesion, and two-stent treatment in 42.0%, 45.9%, and 14.0%, respectively. Proximal optimization technique (POT) or POT-like inflation was performed in 82.1%. GB crossing failure, SB stenting due to dissection, and stent deformation requiring correction by KBI or MV dilation occurred in 8.7, 1.4, and 5.8%, respectively. Finally, simple GB dilation without KBI had been completed in 91.8% for SB dilation. At 1-year follow-up, target lesion revascularization, cardiac death, myocardial infarction, and stent thrombosis were found in 7.2%, 2.1%, 2.1%, and 1.0%, respectively.


CONCLUSION - Simple GB dilation after adequate expansion of proximal MV stent provided acceptable acute and long-term results as an alternative to KBI.