Rotational Atherectomy

Abstract

Recommended Article

Transverse partial stent ablation with rotational atherectomy for suboptimal culotte technique in left main stem bifurcation Rotational atherectomy in the subadventitial space to allow safe and successful chronic total occlusion recanalization: Pushing the limit further Pivotal trial to evaluate the safety and efficacy of the orbital atherectomy system in treating de novo, severely calcified coronary lesions (ORBIT II) Healed coronary plaque rupture as a cause of rapid lesion progression: a case demonstrated with in vivo histopathology by directional coronary atherectomy A Meta-Analysis of Contemporary Lesion Modification Strategies During Percutaneous Coronary Intervention in 244,795 Patients From 22 Studies Trends in Utilization of, and Comparative Safety and Effectiveness of Orbital and Rotational Atherectomy Multicenter Registry of Real-World Patients With Severely Calcified Coronary Lesions Undergoing Orbital Atherectomy: 1-Year Outcomes Five-Year Clinical Outcomes After Drug-Eluting Stent Implantation Following Rotational Atherectomy for Heavily Calcified Lesions

Clinical Trial2013 Oct;6(10):1095-1104.

JOURNAL:JACC Cardiovasc Imaging. Article Link

OCT compared with IVUS in a coronary lesion assessment: the OPUS-CLASS study

Kubo T, Akasaka T, Shite J et al. Keywords: FD; IVUS; MLA; MLD; OCT; PCI; QCA; TD; coronary angiography; frequency domain; intravascular ultrasound; minimum lumen area; minimum lumen diameter; optical coherence tomography; percutaneous coronary intervention; quantitative coronary angiography; time-domain

ABSTRACT


OBJECTIVESThe aim of this study was to investigate the reliability of frequency domain optical coherence tomography (FD-OCT) for coronary measurements compared with quantitative coronary angiography (QCA) and intravascular ultrasound (IVUS).


BACKGROUND - Accurate luminal measurement is expected in FD-OCT because this technology offers high resolution and excellent contrast between lumen and vessel wall.

METHODS - In 5 medical centers, 100 patients with coronary artery disease were prospectively studied by using angiography, FD-OCT, and IVUS. In addition, 5 phantom models of known lumen dimensions (lumen diameter 3.08 mm; lumen area 7.45 mm(2)) were examined using FD-OCT and IVUS. Quantitative image analyses of the coronary arteries and phantom models were performed by an independent core laboratory.

RESULTS - In the clinical study, the mean minimum lumen diameter measured by QCA was significantly smaller than that measured by FD-OCT (1.81 ± 0.72 mm vs. 1.91 ± 0.69 mm; p < 0.001) and the minimum lumen diameter measured by IVUS was significantly greater than that measured by FD-OCT (2.09 ± 0.60 mm vs. 1.91 ± 0.69 mm; p < 0.001). The minimum lumen area measured by IVUS was significantly greater than that by FD-OCT (3.68 ± 2.06 mm(2) vs. 3.27 ± 2.22 mm(2); p < 0.001), although a significant correlation was observed between the 2 imaging techniques (r = 0.95, p < 0.001; mean difference 0.41 mm(2)). Both FD-OCT and IVUS exhibited good interobserver reproducibility, but the root-mean-squared deviation between measurements was approximately twice as high for the IVUS measurements compared with the FD-OCT measurements (0.32 mm(2) vs. 0.16 mm(2)). In a phantom model, the mean lumen area according to FD-OCT was equal to the actual lumen area of the phantom model, with low SD; IVUS overestimated the lumen area and was less reproducible than FD-OCT (8.03 ± 0.58 mm(2) vs. 7.45 ± 0.17 mm(2); p < 0.001).

CONCLUSIONS - The results of this prospective multicenter study demonstrate that FD-OCT provides accurate and reproducible quantitative measurements of coronary dimensions in the clinical setting.

Copyright © 2013 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.