Optical Coherence Tomography-Guided Percutaneous Coronary Intervention in ST-Segment-Elevation Myocardial Infarction: A Prospective Propensity-Matched Cohort of the Thrombectomy Versus Percutaneous Coronary Intervention Alone Trial Optimal threshold of postintervention minimum stent area to predict in-stent restenosis in small coronary arteries: An optical coherence tomography analysis Clinical Impact of Suboptimal Stenting and Residual Intrastent Plaque/Thrombus Protrusion in Patients With Acute Coronary Syndrome: The CLI-OPCI ACS Substudy (Centro per la Lotta Contro L'Infarto-Optimization of Percutaneous Coronary Intervention in Acute Coronary Syndrome) Treatment of calcified coronary lesions with Palmaz-Schatz stents. An intravascular ultrasound study Characteristics of abnormal post-stent optical coherence tomography findings in hemodialysis patients Optical coherence tomography predictors of target vessel myocardial infarction after provisional stenting in patients with coronary bifurcation disease Spontaneous Coronary Artery Dissection: Pathophysiological Insights From Optical Coherence Tomography Japan-United States of America Harmonized Assessment by Randomized Multicentre Study of OrbusNEich's Combo StEnt (Japan-USA HARMONEE) study: primary results of the pivotal registration study of combined endothelial progenitor cell capture and drug-eluting stent in patients with ischaemic coronary disease and non-ST-elevation acute coronary syndrome

Original Research2018 Feb 3;391(10119):431-440.

JOURNAL:Lancet. Article Link

A sirolimus-eluting bioabsorbable polymer-coated stent (MiStent) versus an everolimus-eluting durable polymer stent (Xience) after percutaneous coronary intervention (DESSOLVE III): a randomised, single-blind, multicentre, non-inferiority, phase 3 trial

de Winter RJ, Katagiri Y, Serruys PW et al. Keywords: sirolimus-eluting bioabsorbable polymer-coated stent; everolimus-eluting durable polymer stent;non-inferiority

ABSTRACT

BACKGROUND - MiStent is a drug-eluting stent with a fully absorbable polymer coating containing and embedding a microcrystalline form of sirolimus into the vessel wall. It was developed to overcome the limitation of current durable polymer drug-eluting stents eluting amorphous sirolimus. The clinical effect of MiStent sirolimus-eluting stent compared with a durable polymer drug-eluting stents has not been investigated in a large randomised trial in an all-comer population.

METHODS - We did a randomised, single-blind, multicentre, phase 3 study (DESSOLVE III) at 20 hospitals in Germany, France, Netherlands, and Poland. Eligible participants were any patients aged at least 18 years who underwent percutaneous coronary intervention in a lesion and had a reference vessel diameter of 2·50-3·75 mm. We randomly assigned patients (1:1) to implantation of either a sirolimus-eluting bioresorbable polymer stent (MiStent) or an everolimus-eluting durable polymer stent (Xience). Randomisation was done by local investigators via web-based software with random blocks according to centre. The primary endpoint was a non-inferiority comparison of a device-oriented composite endpoint (DOCE)-cardiac death, target-vessel myocardial infarction, or clinically indicated target lesion revascularisation-between the groups at 12 months after the procedure assessed by intention-to-treat. A margin of 4·0% was defined for non-inferiority of the MiStent group compared with the Xience group. All participants were included in the safety analyses. This trial is registered with ClinicalTrials.gov, number NCT02385279.

FINDINGS - Between March 20, and Dec 3, 2015, we randomly assigned 1398 patients with 2030 lesions; 703 patients with 1037 lesions were assigned to MiStent, of whom 697 received the index procedure, and 695 patients with 993 lesions were asssigned to Xience, of whom 690 received the index procedure. At 12 months, the primary endpoint had occurred in 40 patients (5·8%) in the sirolimus-eluting stent group and in 45 patients (6·5%) in the everolimus-eluting stent group (absolute difference -0·8% [95% CI -3·3 to 1·8], pnon-inferiority=0·0001). Procedural complications occurred in 12 patients (1·7%) in the sirolimus-eluting stent group and ten patients (1·4%) in the everolimus-eluting stent group; no clinical adverse events could be attributed to these dislodgements through a minimum of 12 months of follow-up. The rate of stent thrombosis, a safety indicator, did not differ between groups and was low in both treatment groups.

INTERPRETATION - The sirolimus-eluting bioabsorbable polymer stent was non-inferior to the everolimus-eluting durable polymer stent for a device-oriented composite clinical endpoint at 12 months in an all-comer population. MiStent seems a reasonable alternative to other stents in clinical practice.

FUNDING - The European Cardiovascular Research Institute, Micell Technologies (Durham, NC, USA), and Stentys (Paris, France).

Copyright © 2018 Elsevier Ltd. All rights reserved.