Intravascular optical coherence tomography Angiography Alone Versus Angiography Plus Optical Coherence Tomography to Guide Percutaneous Coronary Intervention Outcomes From the Pan-London PCI Cohort Uncovered Culprit Plaque Ruptures in Patients With ST-Segment Elevation Myocardial Infarction Assessed by Optical Coherence Tomography and Intravascular Ultrasound With iMap Consensus standards for acquisition, measurement, and reporting of intravascular optical coherence tomography studies: a report from the International Working Group for Intravascular Optical Coherence Tomography Standardization and Validation Assessment of the coronary calcification by optical coherence tomography Device specificity of vascular healing following implantation of bioresorbable vascular scaffolds and bioabsorbable polymer metallic drug-eluting stents in human coronary arteries: the ESTROFA OCT BVS vs. BP-DES study Covering our tracks – optical coherence tomography to assess vascular healing Volumetric characterization of human coronary calcification by frequency-domain optical coherence tomography

Clinical TrialVolume 11, Issue 10, May 2018

JOURNAL:JACC Cardiovasc Interv. Article Link

Impact of Optimized Procedure-Related Factors in Drug-Eluting Balloon Angioplasty for Treatment of In-Stent Restenosis

TM Rhee, JM Lee, ES Shin et al. Keywords: drug-eluting balloon; in-stent restenosis; paclitaxel-coated balloonpercutaneous coronary intervention; target lesion failure

ABSTRACT


OBJECTIVES - The aim of this study was to investigate the impact of optimizing procedure-related factors during drug-eluting balloon (DEB) angioplasty on clinical outcomes of drug-eluting stent in-stent restenosis (ISR).


BACKGOURND - Although DEB angioplasty is recommended as a reasonable option for ISR, recurrent target lesion failure (TLF) still occurs in many patients after DEB angioplasty.

METHODES - Consecutive patients with drug-eluting stent ISR treated with DEB (SeQuent Please) were collected from 4 centers in Korea. The primary outcome was 2-year TLF. Procedure-related modifiable independent predictors for TLF and their best cutoff values were determined.

RESULTS - In a total of 256 patients (309 lesions), TLF occurred in 52 patients (20.3%). Modifiable independent predictors of TLF among procedure-related factors were residual diameter stenosis after lesion preparation (residual percentage diameter stenosis [%DS]), DEB-to-stent ratio (BSR), and DEB inflation time (Tinflation), whose best cutoff values were 20%, 0.91, and 60 s, respectively. TLF rates were significantly higher in groups with residual %DS ≥20% (34.7% vs. 12.5%; adjusted hazard ratio: 2.15; 95% confidence interval: 1.86 to 2.48; p < 0.001), BSR ≤0.91 (46.4% vs. 21.9%; adjusted hazard ratio: 2.02; 95% confidence interval: 1.75 to 2.34; p < 0.001), and Tinflation ≤60 s (26.2% vs. 14.0%; adjusted hazard ratio: 1.82; 95% confidence interval: 1.36 to 2.45; p < 0.001). When classifying ISR lesions by combination of procedure-related factors, TLF occurred in 8.3% in the fully optimized procedure group (residual %DS <20%, BSR >0.91, and Tinflation >60 s) and 66.7% in the nonoptimized group (residual %DS ≥20%, BSR ≤0.91, and Tinflation ≤60 s) (p < 0.001).

CONCLUSIONS - Residual %DS after lesion preparation, BSR, and Tinflation were the only modifiable procedure-related factors in DEB angioplasty. Fully optimized DEB angioplasty with optimal lesion preparation, prolonged inflation, and sufficient dilation may play an important role in reducing TLF after DEB angioplasty.