Experimental basis of determining maximum coronary, myocardial, and collateral blood flow by pressure measurements for assessing functional stenosis severity before and after percutaneous transluminal coronary angioplasty Retrospective Comparison of Long-Term Clinical Outcomes Between Percutaneous Coronary Intervention and Medical Therapy in Stable Coronary Artery Disease With Gray Zone Fractional Flow Reserve - COMFORTABLE Retrospective Study Long-term clinical outcome after fractional flow reserve-guided treatment in patients with angiographically equivocal left main coronary artery stenosis Anatomical and Functional Computed Tomography for Diagnosing Hemodynamically Significant Coronary Artery Disease: A Meta-Analysis Coronary Microcirculation Downstream Non-Infarct-Related Arteries in the Subacute Phase of Myocardial Infarction: Implications for Physiology-Guided Revascularization Lesion-Specific and Vessel-Related Determinants of Fractional Flow Reserve Beyond Coronary Artery Stenosis Coronary Flow Reserve in the Instantaneous Wave-Free Ratio/Fractional Flow Reserve Era: Too Valuable to Be Neglected Coronary Physiology in the Cardiac Catheterization Laboratory

Clinical TrialVolume 11, Issue 10, May 2018

JOURNAL:JACC Cardiovasc Interv. Article Link

Comparison of a Novel Biodegradable Polymer Sirolimus-Eluting Stent With a Durable Polymer Everolimus-Eluting Stent 5-Year Outcomes of the Randomized BIOFLOW-II Trial

T Lefèvre, M Haude, FJ Neumann et al. Keywords: biodegradable polymer; coronary artery disease; drug-eluting stent(s); sirolimus

ABSTRACT


OBJECTIVES - The authors aimed to compare long-term data of an ultrathin cobalt-chromium stent with passive silicon carbide coating and an active biodegradable polymer that releases sirolimus (O-SES) (Orsiro, BIOTRONIK, Bülach, Switzerland) with the durable polymer-based Xience Prime everolimus-eluting stent (X-EES) (Abbott Vascular, Santa Clara, California).


BACKGOURND - Biodegradable polymer stents have been developed aiming to overcome long-term detrimental effects of durable polymer stents, ultimately leaving a bare-metal stent in the vessel.

METHODS - This multicenter, assessor-blinded trial randomized 452 patients with 505 lesions to either O-SES or X-EES in a 2:1 fashion. Endpoints at 5 years were target lesion failure (TLF), its components, and stent thrombosis.

RESULTS - TLF occurred in 10.4% (n = 30) of O-SES patients versus 12.7% (n = 19) of X-EES patients (p = 0.473), overall stent thrombosis occurred in 0.7% (n = 2) versus 2.8% (n = 4) (p = 0.088), and definite stent thrombosis in 0% versus 0.7% (n = 1) (p = 0.341). Post hoc analysis was performed in diabetic patients (n = 128) and vessels ≤2.75 mm (n = 259). In diabetic patients, the O-SES group had numerically more target lesion revascularizations (13.5% vs. 4.5%; p = 0.138), but fewer cardiac deaths (1.3% vs. 6.9%; p = 0.089) and stent thrombosis (0% vs. 6.9%; p = 0.039). In small vessels, the O-SES group had a significantly lower 5-year mortality (3.7% vs. 11.3%; p = 0.022).

CONCLUSIONS - At 5 years, the biodegradable polymer O-SES demonstrated low TLF rates comparable to the durable polymer X-EES, confirming its long-term safety and performance. Particularly encouraging is the absence of definite stent thrombosis.