Rotational atherectomy in the subadventitial space to allow safe and successful chronic total occlusion recanalization: Pushing the limit further Prevalence, Presentation and Treatment of 'Balloon Undilatable' Chronic Total Occlusions: Insights from a Multicenter US Registry Rotational atherectomy and new-generation drug-eluting stent implantation Short-term and long-term clinical outcomes of rotational atherectomy in resistant chronic total occlusion Long-term outcomes of rotational atherectomy of underexpanded stents. A single center experience A Meta-Analysis of Contemporary Lesion Modification Strategies During Percutaneous Coronary Intervention in 244,795 Patients From 22 Studies Rotational Atherectomy Followed by Drug-Coated Balloon Dilation for Left Main In-Stent Restenosis in the Setting of Acute Coronary Syndrome Complicated with Right Coronary Chronic Total Occlusion Transverse partial stent ablation with rotational atherectomy for suboptimal culotte technique in left main stem bifurcation

Clinical Trial2004 Jun 2;43(11):1959-63.

JOURNAL:J Am Coll Cardiol. Article Link

Impact of final stent dimensions on long-term results following sirolimus-eluting stent implantation: serial intravascular ultrasound analysis from the sirius trial

Sonoda S, Morino Y, Ako J et al. Keywords: BMS; IVUS; MLA; MSA; SES

ABSTRACT


OBJECTIVES We assessed the predictive value of minimum stent area (MSA) for long-term patency of sirolimus-eluting stents (SES) implantation compared to bare metal stents (BMS).


BACKGROUND - Although MSA is a consistent predictor of in-stent restenosis, its predictive value in BMS is still limited because of biologic variability in the restenosis process.

METHODS - From the SIRolImUS (SIRIUS) trial, 122 cases (SES: 72; BMS: 50) with complete serial intravascular ultrasound (IVUS) (baseline and 8-month follow-up) were analyzed. Postprocedure MSA and follow-up minimum lumen area (MLA) were obtained. Based on previous physiologic studies, adequate stent patency at follow-up was defined as MLA >4 mm(2).

RESULTS - In both groups, a significant positive correlation was observed between baseline MSA and follow-up MLA (SES: p < 0.0001, BMS: p < 0.0001). However, SES showed higher correlation than BMS (0.8 vs. 0.65) with a higher regression coefficient (0.92 vs. 0.59). The sensitivity and specificity curves identified different optimal thresholds of MSA to predict adequate follow-up MLA: 5 mm(2) for SES and 6.5 mm(2) for BMS. The positive predictive values with these cutoff points were 90% and 56%, respectively.

CONCLUSIONS - In this SIRIUS IVUS substudy, SES reduced both biologic variability and restenosis, resulting in increased predictability of long-term stent patency with postprocedure MSA. In addition, SES had a considerably lower optimal MSA threshold compared to BMS.