Percutaneous LAA Occlusion

Abstract

Recommended Article

Does pulsed field ablation regress over time? A quantitative temporal analysis of pulmonary vein isolation Transseptal puncture versus patent foramen ovale or atrial septal defect access for left atrial appendage closure 2015 ACC/HRS/SCAI Left Atrial Appendage Occlusion Device Societal Overview Expert Recommendations on Cardiac Computed Tomography for Planning Transcatheter Left Atrial Appendage Occlusion Percutaneous left atrial appendage occlusion: the Munich consensus document on definitions, endpoints, and data collection requirements for clinical studies Dabigatran dual therapy with ticagrelor or clopidogrel after percutaneous coronary intervention in atrial fibrillation patients with or without acute coronary syndrome: a subgroup analysis from the RE-DUAL PCI trial An artificial intelligence-enabled ECG algorithm for the identification of patients with atrial fibrillation during sinus rhythm: a retrospective analysis of outcome prediction Initial Worldwide Experience With the WATCHMAN Left Atrial Appendage System for Stroke Prevention in Atrial Fibrillation

Clinical Case Study

JOURNAL:CBSMD Article Link

Clinical Case : Significant LM Diameter Mismatch Treated with IVUS-guidance

Bill Gogas, Fei Ye (syndicatedfeeds)

Pre-reading

Procedure

LTRA 6F JL4 GC w/ Seldinger technique.


Buddy wire technique to cross angulated ostium of LAD w/ Runthrough NS.


Following distal placement of BMW wire, Runtrhough NS was retrieved.


Sequential balloon dilatations w/ 2,5 NC Sprinter Legend from distal to proximal with increasing atmospheres from 10 to 18 was performed. IVUS indicated distal LAD w/ RVD of at least 2,5 mm and proximal LAD media to media diameter of 3,5 mm suggested by IVUS (BS, US) which equals (3,5 mm - 0,25 mm) to vessel diameter of 3,25 mm.


Distal LM PB: 45-50% while proximal LM: >55%.


Distal LAD was treated w/ a 2,5 x 35 DES and mid- to prox-LAD up to distal LM with 3.0 x 28 mm DES*.

*DES of 3.0 or 3.5 mm can be post-dilated up to 4.25 mm



Proximal LAD was post-dilated with 3.0 NC Sprinter balloon to reach expansion of 3,5 mm and LM was post-dilated with 4.0 NC balloon to reach diameter of 4.5 mm.


IVUS confirmed MSAs: LM > 11mm2 & LAD > 7 mm2.


Good angiographic result confirmed by IVUS optimal geometric indices post-stent deployments.



Case Presenter Bill Gogas, Interventional Cardiologist, Visiting Professor of Medicine (Cardiology) at Nanjing First Hospital, China

Case Performed in The Spencer B. King III Catheterization Laboratory

Mentors Spencer B. King III, Shao-liang Chen