CBS 2019
CBSMD教育中心
中 文

Other Relevant Articles

Abstract

Recommended Article

Quantitative angiography methods for bifurcation lesions: a consensus statement update from the European Bifurcation Club Burden of 30-Day Readmissions After Percutaneous Coronary Intervention in 833,344 Patients in the United States: Predictors, Causes, and Cost Association Between Living in Food Deserts and Cardiovascular Risk 2017 ESC Guidelines on the Diagnosis and Treatment of Peripheral Arterial Diseases, in collaboration with the European Society for Vascular Surgery (ESVS): Document covering atherosclerotic disease of extracranial carotid and vertebral, mesenteric, renal, upper and lower extremity arteries Endorsed Effects of Aspirin for Primary Prevention in Persons with Diabetes Mellitus Temporal trends in percutaneous coronary interventions thru the drug eluting stent era: Insights from 18,641 procedures performed over 12-year period Blood CSF1 and CXCL12 as Causal Mediators of Coronary Artery Disease Antithrombotic Therapy after Acute Coronary Syndrome or PCI in Atrial Fibrillation

Clinical TrialVolume 13, Issue 1, January 2020

JOURNAL:JACC Cardiovasc Interv. Article Link

Randomized Comparison of Ridaforolimus-Eluting and Zotarolimus-Eluting Coronary Stents 2-Year Clinical Outcomes: From the BIONICS and NIREUS Trials

M Konigstein, PC Smits, MP Love et al. Keywords: PCI; DES; ridaforolimus vs. zotarolimus

ABSTRACT


OBJECTIVES - This study sought to determine clinical outcomes between treatment groups over long-term follow-up.

 

BACKGROUND - The safety and efficacy of a ridaforolimus-eluting stent (RES) was evaluated in the BIONICS (BioNIR Ridaforolimus-Eluting Coronary Stent System in Coronary Stenosis) and NIREUS (BioNIR Ridaforolimus Eluting Coronary Stent System [BioNIR] European Angiography Study) trials, demonstrating noninferiority of RES in comparison with a zotarolimus-eluting stent (ZES) regarding 1-year target lesion failure (TLF) and 6-month angiographic late lumen loss, respectively.

 

METHODS - Patient-level data from the BIONICS (N = 1,919) and NIREUS (N = 302) randomized trials were pooled, and outcomes in patients implanted with RES and ZES compared. Broad inclusion criteria allowed enrollment of patients with acute coronary syndromes and complex lesions. The primary endpoint was the 2-year rate of TLF or clinically driven target lesion revascularization.

 

RESULTS - A total of 2,221 patients (age 63.2 ± 10.3 years; 79.7% men) undergoing percutaneous coronary intervention with RES (n = 1,159) or ZES (n = 1,062) were included. Clinical and angiographic characteristics were similar between groups. At 2 years, the primary endpoint of TLF was similar among patients implanted with RES and ZES (7.0% vs. 7.2%; p = 0.94). Rates of target lesion revascularization (4.8% RES vs. 4.1% ZES; p = 0.41) and target vesselrelated myocardial infarction (3.1% RES vs. 3.8% ZES; p = 0.52) did not differ between groups. The overall rate of stent thrombosis was also similar (0.5% RES vs. 0.9% ZES; p = 0.39).

 

CONCLUSIONS -  In a pooled analysis of 2 randomized trials, 2-year clinical outcomes were similar between patients undergoing percutaneous coronary intervention with RES and ZES. These results support the long-term safety and efficacy of RES for the treatment of a broad population of patients with coronary artery disease.