Incidence and Clinical Outcomes of Stent Fractures on the Basis of 6,555 Patients and 16,482 Drug-Eluting Stents From 4 Centers IVUS in bifurcation stenting: what have we learned? Usefulness of minimum stent cross sectional area as a predictor of angiographic restenosis after primary percutaneous coronary intervention in acute myocardial infarction (from the HORIZONS-AMI Trial IVUS substudy) The relationship between attenuated plaque identified by intravascular ultrasound and no-reflow after stenting in acute myocardial infarction: the HORIZONS-AMI (Harmonizing Outcomes With Revascularization and Stents in Acute Myocardial Infarction) trial Intravascular ultrasound findings of early stent thrombosis after primary percutaneous intervention in acute myocardial infarction: a Harmonizing Outcomes with Revascularization and Stents in Acute Myocardial Infarction (HORIZONS-AMI) substudy Usefulness of intravascular ultrasound to predict outcomes in short-length lesions treated with drug-eluting stents Histopathologic validation of the intravascular ultrasound diagnosis of calcified coronary artery nodules In vivo intravascular ultrasound-derived thin-cap fibroatheroma detection using ultrasound radiofrequency data analysis

Clinical TrialAugust 2018. [Epub ahead]

JOURNAL:JACC Cardiovasc Interv. Article Link

A Randomized Study of Distal Filter Protection Versus Conventional Treatment During Percutaneous Coronary Intervention in Patients With Attenuated Plaque Identified by Intravascular Ultrasound

K Hibi, K Kozuma, S Sonoda et al. Keywords: acute coronary syndrome(s); distal embolism; distal protection device; intravascular ultrasound

ABSTRACT


OBJECTIVES - The aim of this study was to evaluate the utility of distal protection during percutaneous coronary intervention (PCI) in patients with acute coronary syndromes at high risk for distal embolization.


BACKGROUND - The results of previous clinical trials indicated that the routine use of distal protection in patients with ST-segment elevation myocardial infarction did not improve clinical outcomes. However, selective use of distal protection by means of a filter-based distal protection system has not been evaluated.

METHODS - Two hundred patients with acute coronary syndromes who had native coronary artery lesions and attenuated plaque with longitudinal length ≥5 mm on pre-PCI intravascular ultrasound were randomly assigned to undergo PCI with distal protection or conventional treatment.

RESULTS - The primary endpoint (no-reflow phenomenon) occurred in 26 patients (26.5%) in the distal protection group and 39 patients (41.7%) in the conventional treatment group (p = 0.026), and the corrected TIMI (Thrombolysis In Myocardial Infarction) frame count after revascularization was significantly lower in the distal protection group (23 vs. 30.5; p = 0.0003). The incidence of cardiac death, cardiac arrest, cardiogenic shock after revascularization requiring defibrillation, cardiopulmonary resuscitation, or extracorporeal membrane oxygenation was significantly lower in the distal protection group than in the conventional treatment group (0% vs. 5.2%; p = 0.028).

CONCLUSIONS - The use of distal embolic protection applied with a filter device decreased the incidence of the no-reflow phenomenon and was associated with fewer serious adverse cardiac events after revascularization than conventional PCI in patients with acute coronary syndromes with attenuated plaque ≥5 mm in length. (Assessment of Distal Protection Device in Patients at High Risk for Distal Embolism in Acute Coronary Syndrome [ACS] [VAMPIRE3]; NCT01460966)