CBS 2019
CBSMD教育中心
中 文

IVUS Guidance

Abstract

Recommended Article

Mechanisms of in-stent restenosis after drug-eluting stent implantation: intravascular ultrasound analysis Intravascular Ultrasound Guidance vs. Angiographic Guidance in Primary Percutaneous Coronary Intervention for ST-Segment Elevation Myocardial Infarction - Long-Term Clinical Outcomes From the CREDO-Kyoto AMI Registry Intravascular Ultrasound Guidance Reduces Cardiac Death and Coronary Revascularization in Patients Undergoing Drug-Eluting Stent Implantation: Results From a Meta-Analysis of 9 Randomized Trials and 4724 Patients Comparison of intravascular ultrasound versus angiography-guided drug-eluting stent implantation: a meta-analysis of one randomised trial and ten observational studies involving 19,619 patients Impact of plaque components on no-reflow phenomenon after stent deployment in patients with acute coronary syndrome: a virtual histology-intravascular ultrasound analysis Usefulness of intravascular ultrasound to predict outcomes in short-length lesions treated with drug-eluting stents Impact of intravascular ultrasound guidance on long-term mortality in stenting for unprotected left main coronary artery stenosis Differential prognostic effect of intravascular ultrasound use according to implanted stent length

Original Research2009 Aug 1;74(2):348-56.

JOURNAL:Catheter Cardiovasc Interv. Article Link

Defining a new standard for IVUS optimized drug eluting stent implantation: the PRAVIO study

Gerber RT, Latib A, Colombo A et al. Keywords: IVUS guided PCI; angiography-guided PCI; DES; outcome

ABSTRACT


OBJECTIVES - Preliminary Investigation to the Angiographic Versus IVUS Optimization Trial is a single center prospective observational intravascular ultrasound (IVUS) guided stent implantation study assessing new criteria for optimal drug eluting stent (DES) deployment.


BACKGROUND - IVUS assessment of DES often reveals underexpansion and malapposition. Optimal stent deployment is currently poorly defined and previous criteria may not be suitable in long and complex lesions.


METHODS - Optimization was defined as achieving >/or 70% of the cross-sectional area (CSA) of the postdilation balloon. This criterion was applied in 113 complex lesions. The size of this balloon was calculated according to vessel media-to-media diameters at various sites inside the stented segment. The IVUS guided treated lesions were matched according to diabetes, vessel type, reference vessel diameter, minimum lumen diameter (MLD), and lesion length with a group of angiographic treated lesions to compare final MLD achieved.


RESULTS - Mean minimum stent CSA according to the postdilation balloon utilized was 4.62 mm(2), 6.26 mm(2), 7.87 mm(2), and 9.87 mm(2) for 2.5 mm, 3.0 mm, 3.5 mm, and 4 mm balloons, respectively. Final MLD (mm) was significantly larger in the IVUS compared to the angiographic-guided group (3.09 +/- 0.50 vs. 2.67 +/- 0.54; P < 0.0001). There were no procedural complications related to IVUS use.


CONCLUSIONS - We propose new IVUS criteria based on vessel remodeling that results in an increment in the final MLD, compared to angiographic guidance, which is much larger than any previously published study. This criterion seems to be safely achievable. A proposed randomized study (angiographic vs. IVUS optimization trial) has been launched to test these concepts.

 

(c) 2008 Wiley-Liss, Inc.