3-Year Outcomes of the ULTIMATE Trial Comparing Intravascular Ultrasound Versus Angiography-Guided Drug-Eluting Stent Implantation Long-term health outcome and mortality evaluation after invasive coronary treatment using drug eluting stents with or without the IVUS guidance. Randomized control trial. HOME DES IVUS Clinical impact of intravascular ultrasound-guided chronic total occlusion intervention with zotarolimus-eluting versus biolimus-eluting stent implantation: randomized study Intravascular ultrasound-guided unprotected left main coronary artery stenting in the elderly Intravascular ultrasound guidance of percutaneous coronary intervention in ostial chronic total occlusions: a description of the technique and procedural results IVUS Guidance for Coronary Revascularization: When to Start, When to Stop? Effect of Intravascular Ultrasound-Guided vs Angiography-Guided Everolimus-Eluting Stent Implantation: The IVUS-XPL Randomized Clinical Trial Role of intravascular ultrasound in patients with acute myocardial infarction undergoing percutaneous coronary intervention

Original Research2018 Jan;11(1):111-123.

JOURNAL:JACC Cardiovasc Imaging. Article Link

Optical Frequency Domain Imaging Versus Intravascular Ultrasound in Percutaneous Coronary Intervention (OPINION Trial) Results From the OPINION Imaging Study

Otake H, Akasaka T, OPINION Investigators et al. Keywords: intravascular ultrasound; optical coherence tomography; optical frequency domain imaging; percutaneous coronary intervention

ABSTRACT


OBJECTIVES - The authors sought to clarify how intravascular ultrasound (IVUS) and optical coherence tomography affect percutaneous coronary intervention (PCI) with current-generation drug-eluting stents in a pre-specified substudy of the OPINION (OPtical frequency domain imaging versus INtravascular ultrasound in percutaneous coronary interventiON) trial, a multicenter, prospective, randomized, noninferiority trial comparing optical frequency domain imaging (OFDI)-guided PCI with IVUS-guided PCI.


BACKGROUND - The impact of these 2 imaging modalities in guiding PCI remains unknown.


METHODS - Of 829 patients enrolled in the OPINION trial, 106 were included in the present imaging substudy. Their PCI was guided by either IVUS or OFDI, but all patients were imaged by both modalities after PCI and by OFDI at 8 months. Angiographic, OFDI, and IVUS images were analyzed by independent core laboratories, and statistical analysis was done independently by a dedicated institution.


RESULTS -  A total of 103 patients underwent either OFDI-guided (n = 54) or IVUS-guided (n = 49) PCI. Immediately after PCI, OFDI-guided PCI was associated with a smaller trend of minimum stent area (5.28 ± 1.65 mm2 vs. 6.12 ± 2.34 mm2; p = 0.088), fewer proximal stent-edge hematomas (p = 0.04), and fewer irregular protrusions (p = 0.014) than IVUS-guided PCI. At 8 months, the neointima area tended to be smaller in the OFDI-guided PCI group than in the IVUS-guided PCI group (0.56 ± 0.30 mm2 vs. 0.80 ± 0.65 mm2; p = 0.057), although the percentage of uncovered struts was significantly higher in the OFDI-guided PCI group than in the IVUS-guided PCI group (6.97 ± 7.03% vs. 4.67 ± 6.43%; p = 0.039). The minimum lumen area was comparable in both groups (p = 0.18).


CONCLUSIONS - There were several differences in local findings between OFDI- and IVUS-guided PCI as expected given the different protocols for stent sizing in the 2 groups. The minimum lumen area at the 8-month follow-up was comparable, suggesting that OFDI- and IVUS-guided PCI are similarly feasible using the current-generation drug-eluting stents. (OPtical frequency domain imaging versus INtravascular ultrasound in percutaneous coronary interventiON; NCT01873222)