Intravascular ultrasound guidance to minimize the use of iodine contrast in percutaneous coronary intervention: the MOZART (Minimizing cOntrast utiliZation With IVUS Guidance in coRonary angioplasTy) randomized controlled trial Role of Proximal Optimization Technique Guided by Intravascular Ultrasound on Stent Expansion, Stent Symmetry Index, and Side-Branch Hemodynamics in Patients With Coronary Bifurcation Lesions Outcomes with intravascular ultrasound-guided stent implantation: a meta-analysis of randomized trials in the era of drug-eluting stents Mechanical complications of everolimus-eluting stents associated with adverse events: an intravascular ultrasound study Intravascular ultrasound predictors for edge restenosis after newer generation drug-eluting stent implantation Meta-analysis of outcomes after intravascular ultrasound-guided versus angiography-guided drug-eluting stent implantation in 26,503 patients enrolled in three randomized trials and 14 observational studies Consensus from the 5th European Bifurcation Club meeting A three-vessel virtual histology intravascular ultrasound analysis of frequency and distribution of thin-cap fibroatheromas in patients with acute coronary syndrome or stable angina pectoris

Original Research 2018 Sep 30. [Epub ahead of print]

JOURNAL:Catheter Cardiovasc Interv. Article Link

Novel predictor of target vessel revascularization after coronary stent implantation: Intraluminal intensity of blood speckle on intravascular ultrasound

Saito Y, Kitahara H, Okuya Y et al. Keywords: intravascular ultrasound; percutaneous coronary intervention; target vessel revascularization

ABSTRACT


BACKGROUND - The difference in intraluminal intensity of blood speckle (IBS) on integrated backscatter-intravascular ultrasound (IB-IVUS) across the coronary artery stenosis (i.e., ΔIBS) has been reported to negatively correlate with fractional flow reserve. Fractional flow reserve after coronary stenting is known as a predictor of target vessel revascularization (TVR). However, the relation between ΔIBS and TVR is unclear.


METHODS - Seven hundred and three vessels which underwent percutaneous coronary intervention with stents were screened. Vessels without IVUS-guidance and follow-up information were excluded. Intraluminal IBS values were measured using IB-IVUS in cross-sections at the ostium of the target vessel and at the distal reference of implanted stent. ΔIBS was calculated as (distal IBS) - (ostium IBS).


RESULTS - A total of 393 vessels were included. Mean ΔIBS at postprocedure was 6.22 ± 5.65. During the follow-up period (11.2 ± 3.1 months), 24 cases (6.1%) had TVR. ΔIBS was significantly greater in the vessels with TVR than in those without (11.10 ± 5.93 vs. 5.90 ± 5.49, P <0.001). In receiver operating characteristic curve analysis, ΔIBS significantly predicted TVR (AUC 0.74, best cut-off value 8.24, P < 0.001). Multiple logistic regression analysis showed use of drug eluting stent and ΔIBS ≥ 8.24 as independent predictors of TVR.


CONCLUSIONS - ΔIBS at postprocedure was significantly associated with TVR. IVUS may be able to predict TVR by physiological assessment with measurement of ΔIBS.


© 2018 Wiley Periodicals, Inc.