Temporal Trends in Inpatient Use of Intravascular Imaging Among Patients Undergoing Percutaneous Coronary Intervention in the United States Comparison of intravascular ultrasound guided versus angiography guided drug eluting stent implantation: a systematic review and meta-analysis Impact of intravascular ultrasound-guided percutaneous coronary intervention on long-term clinical outcomes in a real world population First-in-man evaluation of intravascular optical frequency domain imaging (OFDI) of Terumo: a comparison with intravascular ultrasound and quantitative coronary angiography Relationship between intravascular ultrasound guidance and clinical outcomes after drug-eluting stents: the assessment of dual antiplatelet therapy with drug-eluting stents (ADAPT-DES) study Assessment of coronary atherosclerosis by IVUS and IVUS-based imaging modalities: progression and regression studies, tissue composition and beyond Intravascular ultrasound-guided implantation of drug-eluting stents to improve outcome: a meta-analysis Clinical impact of intravascular ultrasound-guided chronic total occlusion intervention with zotarolimus-eluting versus biolimus-eluting stent implantation: randomized study

Clinical Trial2004 Jun 2;43(11):1959-63.

JOURNAL:J Am Coll Cardiol. Article Link

Impact of final stent dimensions on long-term results following sirolimus-eluting stent implantation: serial intravascular ultrasound analysis from the sirius trial

Sonoda S, Morino Y, Ako J et al. Keywords: BMS; IVUS; MLA; MSA; SES

ABSTRACT


OBJECTIVES We assessed the predictive value of minimum stent area (MSA) for long-term patency of sirolimus-eluting stents (SES) implantation compared to bare metal stents (BMS).


BACKGROUND - Although MSA is a consistent predictor of in-stent restenosis, its predictive value in BMS is still limited because of biologic variability in the restenosis process.

METHODS - From the SIRolImUS (SIRIUS) trial, 122 cases (SES: 72; BMS: 50) with complete serial intravascular ultrasound (IVUS) (baseline and 8-month follow-up) were analyzed. Postprocedure MSA and follow-up minimum lumen area (MLA) were obtained. Based on previous physiologic studies, adequate stent patency at follow-up was defined as MLA >4 mm(2).

RESULTS - In both groups, a significant positive correlation was observed between baseline MSA and follow-up MLA (SES: p < 0.0001, BMS: p < 0.0001). However, SES showed higher correlation than BMS (0.8 vs. 0.65) with a higher regression coefficient (0.92 vs. 0.59). The sensitivity and specificity curves identified different optimal thresholds of MSA to predict adequate follow-up MLA: 5 mm(2) for SES and 6.5 mm(2) for BMS. The positive predictive values with these cutoff points were 90% and 56%, respectively.

CONCLUSIONS - In this SIRIUS IVUS substudy, SES reduced both biologic variability and restenosis, resulting in increased predictability of long-term stent patency with postprocedure MSA. In addition, SES had a considerably lower optimal MSA threshold compared to BMS.