Refractory Angina: From Pathophysiology to New Therapeutic Nonpharmacological Technologies Coronary Catheterization and Percutaneous Coronary Intervention in China: 10-Year Results From the China PEACE-Retrospective CathPCI Study A systematic review of factors predicting door to balloon time in ST-segment elevation myocardial infarction treated with percutaneous intervention China PEACE risk estimation tool for in-hospital death from acute myocardial infarction: an early risk classification tree for decisions about fibrinolytic therapy Trends and Impact of Door-to-Balloon Time on Clinical Outcomes in Patients Aged <75, 75 to 84, and ≥85 Years With ST-Elevation Myocardial Infarction Fine particulate air pollution and hospital admissions and readmissions for acute myocardial infarction in 26 Chinese cities Symptom-Onset-To-Balloon Time, ST-Segment Resolution and In-Hospital Mortality in Patients With ST-Segment Elevation Myocardial Infarction Undergoing Primary Percutaneous Coronary Intervention in China: From China Acute Myocardial Infarction Registry Natural History of Spontaneous Coronary Artery Dissection With Spontaneous Angiographic Healing

Clinical Trial2018 Aug 16.

JOURNAL:Cardiovasc Revasc Med. Article Link

A randomised trial comparing two stent sizing strategies in coronary bifurcation treatment with bioresorbable vascular scaffolds - The Absorb Bifurcation Coronary (ABC) trial

Rampat R, Mayo T, Hildick-Smith D et al. Keywords: Absorb; BVS; Bioresorbable vascular scaffold; Coronary bifurcation

ABSTRACT


BACKGROUND - Limited information is available on the use of Bioresorbable Vascular Scaffold (BVS) in bifurcations involving significant side branches. When treating bifurcation disease with metal stents, the recommendation is to choose a stent diameter based on the distal main vessel diameter. Whether this sizing strategy is applicable to BVS is currently unknown.


METHODS - We randomised 37 patients undergoing elective PCI for 'false' bifurcation disease (Medina 0,1,0; 1,0,0; 1,1,0) to receive BVS based either on proximal or distal reference diameters. Optical Frequency Domain Imaging (OFDI) measurements were performed pre BVS insertion to obtain proximal and distal reference diameters and post implantation. BVS size was chosen according to the proximal or distal reference diameter as per randomisation. Implantation was performed using the PSP technique tailored to bifurcation stenting. OFDI was repeated post implantation to confirm satisfactory expansion and apposition.

RESULTS - Baseline demographics between the two groups were similar. Patients were aged 62.8 ± 3.3 years; 76% were male. Mean side branch diameter was 2.24 ± 0.13 mm. TIMI III flow in the main vessel was achieved in all cases. Side branch occlusion occurred in 1 case (2.7%). In the distal-sizing arm, there was a greater incidence of significant malapposition (>300 μm) at the proximal end of the scaffold on OCT (2.3% versus 0.8%, p 0.023). The incidence of distal edge dissections was numerically greater in the proximal-sizing group but this was not statistically significant (31.3% vs 11.8%, p 0.17).

CONCLUSION - Both proximal and distal sizing strategies have similar procedural complication rates when using the ABSORB BVS to treat coronary bifurcations. However a proximal sizing strategy is associated with less malapposition and may be preferable.