CBS 2019
CBSMD教育中心
中 文

Optical Coherence Tomography

Abstract

Recommended Article

Histopathological validation of optical coherence tomography findings of the coronary arteries Device specificity of vascular healing following implantation of bioresorbable vascular scaffolds and bioabsorbable polymer metallic drug-eluting stents in human coronary arteries: the ESTROFA OCT BVS vs. BP-DES study Clinical Impact of OCT Findings During PCI: The CLI-OPCI II Study Superficial Calcium Fracture After PCI as Assessed by OCT Neoatherosclerosis in Patients With Coronary Stent Thrombosis: Findings From Optical Coherence Tomography Imaging (A Report of the PRESTIGE Consortium) Characteristics of stent thrombosis in bifurcation lesions analysed by optical coherence tomography Lipid-rich plaque and myocardial perfusion after successful stenting in patients with non-ST-segment elevation acute coronary syndrome: an optical coherence tomography study Optical Coherence Tomography to Assess Proximal Side Optimization Technique in Crush Stenting

Review Article2015 Jul;4(3):285-294.

JOURNAL:Interv Cardiol Clin. Article Link

Assessment and Quantitation of Stent Results by Intracoronary Optical Coherence Tomography

Maehara A, Matsumura M, Mintz GS. Keywords: dissection; malapposition; OCT; stent; minimal stent area

ABSTRACT

Optical coherence tomography evaluation of poststent results includes stent expansion as the absolute minimum stent area ratio by comparing the minimum stent area with the proximal and distal reference lumen areas or mean stent area defined as the total stent volume divided by the analyzed stent length; stent strut malapposition defined when the distance from the center of the blooming artifact and the surface of plaque is greater than the sum of stent thickness and polymer thickness; tissue protrusion through the stent struts; semiquantitative residual thrombus evaluation; and stent edge dissection.