Fractional Flow Reserve

Abstract

Recommended Article

Clinical value of post-percutaneous coronary intervention fractional flow reserve value: A systematic review and meta-analysis Comparison of Accuracy of One-Use Methods for Calculating Fractional Flow Reserve by Intravascular Optical Coherence Tomography to That Determined by the Pressure-Wire Method Fractional Flow Reserve-Guided Multivessel Angioplasty in Myocardial Infarction Clinical Significance of Concordance or Discordance Between Fractional Flow Reserve and Coronary Flow Reserve for Coronary Physiological Indices, Microvascular Resistance, and Prognosis After Elective Percutaneous Coronary Intervention Diagnostic performance of stress perfusion cardiac magnetic resonance for the detection of coronary artery disease: A systematic review and meta-analysis Instantaneous Wave-free Ratio versus Fractional Flow Reserve to Guide PCI Prognostic Implication of Thermodilution Coronary Flow Reserve in Patients Undergoing Fractional Flow Reserve Measurement Coronary Flow Reserve in the Instantaneous Wave-Free Ratio/Fractional Flow Reserve Era: Too Valuable to Be Neglected

Original Research2017 Dec;30(6):564-569.

JOURNAL:J Interv Cardiol. Article Link

Diagnostic accuracy of instantaneous wave free-ratio in clinical practice

Ding WY, Nair S, Appleby C. Keywords: fractional flow reserve; functional testing; instantaneous wave-free ratio; pressure wire studies

ABSTRACT


AIMS - To evaluate the correlation between iFR and FFR in real-world clinical practice.


METHODS AND RESULTS - Retrospective, single-centre study of 229 consecutive pressure-wire studies (np  = 158). Real-time iFR and FFR measurements were performed for angiographically borderline stenoses. Functionally significant stenoses were defined as iFR <0.86 or FFR ≤0.80. An iFR between 0.86 and 0.93 was considered within the grey zone (Hybrid approach). Median iFR and FFR (IQR) were 0.92 (0.87-0.95) and 0.83 (0.76-0.89), respectively. Pearson's correlation coefficient was 0.75 (P < 0.001). Bland-Altman plot showed a mean difference between iFR and FFR that remained consistent throughout the range of values. The optimal iFR cutoff was 0.91-sensitivity 80%, specificity 82% with ROC area under curve of 89%. Using the Hybrid iFR-FFR strategy, we demonstrated high accuracy of iFR results-sensitivity 95%, specificity 96%, PPV 95%, and NPV 96%. In addition, this method would have avoided adenosine in 56% of patients. Mean follow-up period was 17.2 (±3.4) months. All-cause mortality was 3.2% (np = 5) and repeat intervention was required in six lesions (2.6%).


CONCLUSIONS - This study demonstrates that iFR is a valuable adjunct to FFR using the Hybrid iFR-FFR strategy in a real-world population. The use of adenosine may be avoided in about half the cases.


© 2017, Wiley Periodicals, Inc.