Recommended Article

Active SB-P Versus Conventional Approach to the Protection of High-Risk Side Branches: The CIT-RESOLVE Trial Validation of bifurcation DEFINITION criteria and comparison of stenting strategies in true left main bifurcation lesions Classification and treatment of coronary artery bifurcation lesions: putting the Medina classification to the test T and small protrusion (TAP) vs double kissing crush technique: Insights from in-vitro models One Versus 2-stent Strategy for the Treatment of Bifurcation Lesions in the Context of a Coronary Chronic Total Occlusion: A Multicenter Registry Optimal Fluoroscopic Projections of Coronary Ostia and Bifurcations Defined by Computed Tomographic Coronary Angiography ‘Small bifurcation?’ CT myocardial mass volume measurements change therapeutic strategy in coronary artery disease Randomized study on simple versus complex stenting of coronary artery bifurcation lesions: the Nordic bifurcation study

Original Research2020 Jul 14;EIJ-D-20-00334.

JOURNAL:Eurointervention. Article Link

Feasibility and efficacy of the ultrashort side branch dedicated balloon in coronary bifurcation stenting

Y Murasato, M Nishihara, T Mori et al. Keywords: kissing balloon inflation vs. glider balloon; stent configuration; high-angled bifurcation; coronary bifurcation stenting

ABSTRACT

AIMS - We sought to investigate the feasibility and efficacy of the Glider balloon (GB), a side branch (SB) dedicated balloon 4-mm in length, in coronary bifurcation stenting.

METHODS AND RESULTS - In bifurcation bench models, stent configuration was examined with micro-focus computed tomography after crossover stenting followed by GB dilation or kissing balloon inflation (KBI). GB dilation maintained cross-sectional stent area without significant deformation and presented effective jailed strut removal in a high-angled bifurcation model.


We performed GB dilatation after main vessel (MV) stenting for 207 lesions in 194 patients, which included left main, true-bifurcation lesion, and two-stent treatment in 42.0%, 45.9%, and 14.0%, respectively. Proximal optimization technique (POT) or POT-like inflation was performed in 82.1%. GB crossing failure, SB stenting due to dissection, and stent deformation requiring correction by KBI or MV dilation occurred in 8.7, 1.4, and 5.8%, respectively. Finally, simple GB dilation without KBI had been completed in 91.8% for SB dilation. At 1-year follow-up, target lesion revascularization, cardiac death, myocardial infarction, and stent thrombosis were found in 7.2%, 2.1%, 2.1%, and 1.0%, respectively.


CONCLUSION - Simple GB dilation after adequate expansion of proximal MV stent provided acceptable acute and long-term results as an alternative to KBI.