Three-Year Outcomes of the DKCRUSH-V Trial Comparing DK Crush With Provisional Stenting for Left Main Bifurcation Lesions Double-Kissing Culotte Technique for Coronary Bifurcation Stenting - Technical evaluation and comparison with conventional double stenting techniques A Randomized Trial Evaluating Online 3-Dimensional Optical Frequency Domain Imaging-Guided Percutaneous Coronary Intervention in Bifurcation Lesions Metabolic Interactions and Differences between Coronary Heart Disease and Diabetes Mellitus: A Pilot Study on Biomarker Determination and Pathogenesis Clinical Outcome of Double Kissing Crush Versus Provisional Stenting of Coronary Artery Bifurcation Lesions: The 5-Year Follow-Up Results From a Randomized and Multicenter DKCRUSH-II Study (Randomized Study on Double Kissing Crush Technique Versus Provisional Stenting Technique for Coronary Artery Bifurcation Lesions) Feasibility and efficacy of the ultrashort side branch dedicated balloon in coronary bifurcation stenting Randomized study of the crush technique versus provisional side-branch stenting in true coronary bifurcations: the CACTUS (Coronary Bifurcations: Application of the Crushing Technique Using Sirolimus-Eluting Stents) Study Tips of the dual-lumen microcatheter-facilitated reverse wire technique in percutaneous coronary interventions for markedly angulated bifurcated lesions

Original ResearchAvailable online 10 September 2020

JOURNAL:Cardiovasc Revasc Med. Article Link

T and small protrusion (TAP) vs double kissing crush technique: Insights from in-vitro models

V Paradies, Jaryl Ng, HY Ang et al. Keywords: bifurcation lesions; double stenting; in-vitro models; thrombogenicity

ABSTRACT


BACKGROUND - Percutaneous coronary interventions on complex bifurcation lesions may require implantation of two stents to appropriately treat diffuse side-branch (SB) disease. Comparisons among different bifurcation stenting techniques are continuously attempted by various study designs (bench tests, computer simulations, clinical studies). Among different techniques, double kissing crush (DKC) represents the last evolution for crushingwhile T and small Protrusion (TAP) represents the evolution of T stenting. Both techniques are actually gaining popularity, but head-to-head comparisons are lacking.

 

METHODS AND RESULTS - Two last generation drug-eluting stents (Synergy, Boston Scientific, MA, USA and Ultimaster,Terumo Corp., Japan) were implanted in left main bifurcation bench models using TAP (n = 6 sets) and DKC (n = 6 sets) techniques. A peristaltic pump with fresh porcine blood was used to perfuse the blood through the silicone model at a flow rate of 200 ml/min for 4 min. Optical coherence tomography (OCT) was used to assess stent struts geometry and thrombus formation. SB cross sectional area as well as SB obstruction did not significantly differ between the two techniques. Numerical (but not statistically significant) differences were found in terms of malapposed struts (fewer with TAP) and floating struts (fewer with DKC). Thrombus formation after blood perfusion was similar between TAP and DKC technique (1.53 ± 1.12 vs. 1.20 ± 1.01 mm2, p = 0.6).

 

CONCLUSION - The result of the present in-vitro study shows the absence of significant difference between TAP and DKC in terms of stent struts apposition and acute thrombus formation potential. Despite the completely different technical steps required, both techniques have similar performance according to such articulated pre-clinical evaluation.