Developing a Mobile Application for Global Cardiovascular Education PCI for obstructive bifurcation lesions the 14th consensus document from the european bifurcation club Double-Kiss-Crush Bifurcation Stenting: Step-by-Step Troubleshooting Streamlined reverse wire technique for the treatment of complex bifurcated lesions The EBC TWO Study (European Bifurcation Coronary TWO): A Randomized Comparison of Provisional T-Stenting Versus a Systematic 2 Stent Culotte Strategy in Large Caliber True Bifurcations Comparison of intravascular ultrasound-guided with angiography-guided double kissing crush stenting for patients with complex coronary bifurcation lesions: rationale and design of a prospective, randomized and multicenter DKCRUSH VIII trial Technical aspects of the culotte technique Incidence of Adverse Events at 3 Months Versus at 12 Months After Dual Antiplatelet Therapy Cessation in Patients Treated With Thin Stents With Unprotected Left Main or Coronary Bifurcations

Original ResearchAvailable online 10 September 2020

JOURNAL:Cardiovasc Revasc Med. Article Link

T and small protrusion (TAP) vs double kissing crush technique: Insights from in-vitro models

V Paradies, Jaryl Ng, HY Ang et al. Keywords: bifurcation lesions; double stenting; in-vitro models; thrombogenicity

ABSTRACT


BACKGROUND - Percutaneous coronary interventions on complex bifurcation lesions may require implantation of two stents to appropriately treat diffuse side-branch (SB) disease. Comparisons among different bifurcation stenting techniques are continuously attempted by various study designs (bench tests, computer simulations, clinical studies). Among different techniques, double kissing crush (DKC) represents the last evolution for crushingwhile T and small Protrusion (TAP) represents the evolution of T stenting. Both techniques are actually gaining popularity, but head-to-head comparisons are lacking.

 

METHODS AND RESULTS - Two last generation drug-eluting stents (Synergy, Boston Scientific, MA, USA and Ultimaster,Terumo Corp., Japan) were implanted in left main bifurcation bench models using TAP (n = 6 sets) and DKC (n = 6 sets) techniques. A peristaltic pump with fresh porcine blood was used to perfuse the blood through the silicone model at a flow rate of 200 ml/min for 4 min. Optical coherence tomography (OCT) was used to assess stent struts geometry and thrombus formation. SB cross sectional area as well as SB obstruction did not significantly differ between the two techniques. Numerical (but not statistically significant) differences were found in terms of malapposed struts (fewer with TAP) and floating struts (fewer with DKC). Thrombus formation after blood perfusion was similar between TAP and DKC technique (1.53 ± 1.12 vs. 1.20 ± 1.01 mm2, p = 0.6).

 

CONCLUSION - The result of the present in-vitro study shows the absence of significant difference between TAP and DKC in terms of stent struts apposition and acute thrombus formation potential. Despite the completely different technical steps required, both techniques have similar performance according to such articulated pre-clinical evaluation.