CBS 2019
CBSMD教育中心
中 文

经导管主动脉瓣置换

Abstract

Recommended Article

Balloon Aortic Valvuloplasty as a Bridge to Aortic Valve Replacement: A Contemporary Nationwide Perspective Rationale and design of a randomized clinical trial comparing safety and efficacy of Myval transcatheter heart valve versus contemporary transcatheter heart valves in patients with severe symptomatic aortic valve stenosis: the LANDMARK trial 2015 ESC Guidelines for the management of infective endocarditis: The Task Force for the Management of Infective Endocarditis of the European Society of Cardiology (ESC) Endorsed by: European Association for Cardio-Thoracic Surgery (EACTS), the European Association of Nuclear Medicine (EANM) Risk of Coronary Obstruction and Feasibility of Coronary Access After Repeat Transcatheter Aortic Valve Replacement With the Self-Expanding Evolut Valve: A Computed Tomography Simulation Study Comparison of newer generation self-expandable vs. balloon-expandable valves in transcatheter aortic valve implantation: the randomized SOLVE-TAVI trial Left Ventricular Hypertrophy and Clinical Outcomes Over 5 Years After TAVR: An Analysis of the PARTNER Trials and Registries Comparison of Early Surgical or Transcatheter Aortic Valve Replacement Versus Conservative Management in Low-Flow, Low-Gradient Aortic Stenosis Using Inverse Probability of Treatment Weighting: Results From the TOPAS Prospective Observational Cohort Study Infective endocarditis after transcatheter aortic valve implantation: a nationwide study

Original ResearchVolume 13, Issue 6, March 2020

JOURNAL:JACC Cardiovasc Interv. Article Link

Coronary Access After TAVR With a Self-Expanding Bioprosthesis: Insights From Computed Tomography

M Abdelghani, M Landt, H Traboulsi et al. Keywords: access; angiography; coronary intervention; transcatheter aortic valve

ABSTRACT


OBJECTIVES - The authors sought to estimate possible interference of the Medtronic Evolut R/Pro transcatheter heart valve (THV) frame with coronary access using multislice computed tomography (MSCT) data.

 

BACKGROUND - Lower-risk patients undergoing transcatheter aortic valve replacement (TAVR) endure a high cumulative risk of coronary events, but coronary access can be challenging.

 

METHODS - In 101 patients who received an Evolut R/Pro THV, post-TAVR MSCT (performed at a median of 30 days after TAVR) was used to assess possible interference of the elements of the THV frame with coronary access.

 

RESULTS - The closest cell of the THV frame vertically aligned with the coronary ostium was located opposite the ostium in 58% and 63%, below the ostium in 22% and 30%, or above the ostium in 20% and 7% of left and right coronary arteries, respectively. The free sinus of Valsalva space between the THV frame and the coronary ostium was 0.45 ± 0.17 cm and 0.44 ± 0.17 cm for the left and right coronary arteries, respectively, and showed a stepwise decrease with decreasing THV size (p < 0.001). Bioprosthetic valve commissures were antianatomic (i.e., not aligned with native commissures) in 45 patients (47%), and the commissural post was overlapping a coronary ostium in 15 patients (16%). Two patients (2.0%) had a possible interference of the paravalvular sealing skirt with coronary access.

 

CONCLUSIONS - Using post-TAVR MSCT data, the main mechanism of potential interference of Evolut R/Pro frame with coronary access was an antianatomic commissural post overlapping the coronary ostium.