CBS 2019
CBSMD教育中心
中 文

Transcatheter Aortic Valve Replacement

Abstract

Recommended Article

Aortic Valve Stenosis Treatment Disparities in the Underserved JACC Council Perspectives Online Quantitative Aortographic Assessment of Aortic Regurgitation After TAVR: Results of the OVAL Study Impact of Pre-Existing and New-Onset Atrial Fibrillation on Outcomes After Transcatheter Aortic Valve Replacement Left Ventricular Rapid Pacing Via the Valve Delivery Guidewire in Transcatheter Aortic Valve Replacement Long-term outcome of prosthesis-patient mismatch after transcatheter aortic valve replacement Acute Aortic Syndrome Revisited: JACC State-of-the-Art Review Prognostic Value of Computed Tomography-Derived Extracellular Volume in TAVR Patients With Low-Flow Low-Gradient Aortic Stenosis Frailty and Bleeding in Older Adults Undergoing TAVR or SAVR: Insights From the FRAILTY-AVR Study

Clinical TrialVolume 13, Issue 5, March 2020

JOURNAL:JACC Cardiovasc Interv. Article Link

Prior Balloon Valvuloplasty Versus Direct Transcatheter Aortic Valve Replacement: Results From the DIRECTAVI Trial

F Leclercq, P Robert, M Akodad et al. Keywords: balloon aortic valvuloplasty vs TAVR; device success; direct implantation

ABSTRACT


OBJECTIVES - The aim of this study was to evaluate device success of transcatheter aortic valve replacement (TAVR) using new-generation balloon-expandable prostheses with or without balloon aortic valvuloplasty (BAV).

 

BACKGROUND - Randomized studies are lacking comparing TAVR without BAV against the conventional technique of TAVR with BAV.

 

METHODS - DIRECTAVI (Direct Transcatheter Aortic Valve Implantation) was an open-label noninferiority study that randomized patients undergoing TAVR using the Edwards SAPIEN 3 valve with or without prior balloon valvuloplasty. The primary endpoint was the device success rate according to Valve Academic Research Consortium-2 criteria, which was evaluated using a 7% noninferiority margin. The secondary endpoint included procedural and 30-day adverse events.

 

RESULTS - Device success was recorded for 184 of 236 included patients (78.0%). The rate of device success in the direct implantation group (n = 97 [80.2%]) was noninferior to that in the BAV group (n = 87 [75.7%]) (mean difference 4.5%; 95% confidence interval: ?4.4% to 13.4%; p = 0.02 for noninferiority). No severe prosthesis-patient mismatch or severe aortic regurgitation occurred in any group. In the direct implantation group, 7 patients (5.8%) required BAV to cross the valve. Adverse events were related mainly to pacemaker implantation (20.9% in the BAV group vs. 19.0% in the direct implantation group; p = 0.70). No significant difference was found between the 2 strategies in duration of procedure, contrast volume, radiation exposure, or rate of post-dilatation.

 

CONCLUSIONS - Direct TAVR without prior BAV was noninferior to the conventional strategy using BAV with new-generation balloon-expandable valves, but without procedural simplification. BAV was needed to cross the valve in a few patients, suggesting a need for upstream selection on the basis of patient anatomy. (TAVI Without Balloon Predilatation [of the Aortic Valve] SAPIEN 3 [DIRECTAVI]; NCT02729519)