CBS 2019
CBSMD教育中心
中 文

Transcatheter Aortic Valve Replacement

Abstract

Recommended Article

Extracellular Myocardial Volume in Patients With Aortic Stenosis Prognostic Value of Computed Tomography-Derived Extracellular Volume in TAVR Patients With Low-Flow Low-Gradient Aortic Stenosis Transcatheter and surgical aortic valve replacement in patients with bicuspid aortic valve 2020 ACC Expert Consensus Decision Pathway on Management of Conduction Disturbances in Patients Undergoing Transcatheter Aortic Valve Replacement A Report of the American College of Cardiology Solution Set Oversight Committee Transcatheter Aortic Valve Replacement in Low-risk Patients With Bicuspid Aortic Valve Stenosis Leaflet immobility and thrombosis in transcatheter aortic valve replacement Timing of intervention in asymptomatic patients with valvular heart disease Considerations for Optimal Device Selection in Transcatheter Aortic Valve Replacement: A Review

Original ResearchVolume 13, Issue 6, March 2020

JOURNAL:JACC Cardiovasc Interv. Article Link

Coronary Access After TAVR With a Self-Expanding Bioprosthesis: Insights From Computed Tomography

M Abdelghani, M Landt, H Traboulsi et al. Keywords: access; angiography; coronary intervention; transcatheter aortic valve

ABSTRACT


OBJECTIVES - The authors sought to estimate possible interference of the Medtronic Evolut R/Pro transcatheter heart valve (THV) frame with coronary access using multislice computed tomography (MSCT) data.

 

BACKGROUND - Lower-risk patients undergoing transcatheter aortic valve replacement (TAVR) endure a high cumulative risk of coronary events, but coronary access can be challenging.

 

METHODS - In 101 patients who received an Evolut R/Pro THV, post-TAVR MSCT (performed at a median of 30 days after TAVR) was used to assess possible interference of the elements of the THV frame with coronary access.

 

RESULTS - The closest cell of the THV frame vertically aligned with the coronary ostium was located opposite the ostium in 58% and 63%, below the ostium in 22% and 30%, or above the ostium in 20% and 7% of left and right coronary arteries, respectively. The free sinus of Valsalva space between the THV frame and the coronary ostium was 0.45 ± 0.17 cm and 0.44 ± 0.17 cm for the left and right coronary arteries, respectively, and showed a stepwise decrease with decreasing THV size (p < 0.001). Bioprosthetic valve commissures were antianatomic (i.e., not aligned with native commissures) in 45 patients (47%), and the commissural post was overlapping a coronary ostium in 15 patients (16%). Two patients (2.0%) had a possible interference of the paravalvular sealing skirt with coronary access.

 

CONCLUSIONS - Using post-TAVR MSCT data, the main mechanism of potential interference of Evolut R/Pro frame with coronary access was an antianatomic commissural post overlapping the coronary ostium.