CBS 2019
CBSMD教育中心
中 文

Scientific Library

Abstract

Recommended Article

The SABRE Trial (Sirolimus Angioplasty Balloon for Coronary In-Stent Restenosis): Angiographic Results and 1-Year Clinical Outcomes Trial Design Principles for Patients at High Bleeding Risk Undergoing PCI: JACC Scientific Expert Panel New AHA/ACC/HRS Guidance on Sudden Cardiac Death Prevention The optimal duration of dual antiplatelet therapy after coronary stent implantation: to go too far is as bad as to fall short Incidence and Clinical Outcomes of Stent Fractures on the Basis of 6,555 Patients and 16,482 Drug-Eluting Stents From 4 Centers Biological Phenotypes of Heart Failure With Preserved Ejection Fraction Outcomes After Left Main Percutaneous Coronary Intervention Versus Coronary Artery Bypass Grafting According to Lesion Site: Results From the EXCEL Trial Effect of a Home-Based Wearable Continuous ECG Monitoring Patch on Detection of Undiagnosed Atrial Fibrillation The mSToPS Randomized Clinical Trial

Clinical Trial2017 Oct 23;10(20):2029-2037

JOURNAL:JACC Cardiovasc Interv. Article Link

The SABRE Trial (Sirolimus Angioplasty Balloon for Coronary In-Stent Restenosis): Angiographic Results and 1-Year Clinical Outcomes

Verheye S, Vrolix M, Kumsars I et al. Keywords: drug-eluting balloon; extended release; in-stent restenosis; porous angioplasty balloon; sirolimus nanoparticle

ABSTRACT



OBJECTIVES - The aim of this first-in-human study was to assess the safety and effectiveness of the Virtue sirolimus-eluting balloon in a cohort of patients with in-stent restenosis (ISR).


BACKGROUND - Angioplasty balloons coated with the cytotoxic drug paclitaxel have been widely used for ISR treatment. The Virtue angioplasty balloon (Caliber Therapeutics, New Hope, Pennsylvania) delivers sirolimus in a nanoencapsulated liquid formulation. This clinical trial is the first to examine a sirolimus-eluting balloon for ISR.


METHODS - In this prospective, single-arm feasibility study at 9 European centers, 50 ISR patients were treated with the Virtue balloon. Angiographic measurements at 6 months are reported, along with 12-month clinicalfollow-up.


RESULTS - Procedural success in the intention-to-treat population was 100%. The primary safety endpoint was target lesion failure (TLF) (cardiac death, target vessel myocardial infarction, and clinically driven target lesion revascularization) assessed at 30 days (0%, n = 50). The primary performance endpoint was in-segment late lumen loss (LLL) at 6 months (0.31 ± 0.52 mm; n = 47). Secondary 6-month endpoints include binary restenosis (19.1%), diameter stenosis (30.3 ± 19.9%), and major adverse cardiac events (MACE) (10.2%, n = 49). In the 36-patient per-protocol population (excluding major protocol violations and previously stented ISR), LLL was 0.12 ± 0.33 mm at 6 months. Clinical outcomes at 1 year for the intention-to-treat group were 12.2% TLF and 14.3% MACE and for the per-protocol population were 2.8% TLF and 2.8% MACE.


CONCLUSIONS - This first-in-human study showed excellent procedural success for the Virtue sirolimus-eluting angioplasty balloon, 6-month LLL rates in line with current stent-free ISR treatment options, and clinicaloutcomes that warrant further evaluation in dedicated randomized studies.


Copyright © 2017 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.