Use of Intravascular Ultrasound Imaging in Percutaneous Coronary Intervention to Treat Left Main Coronary Artery Disease Attenuated plaque detected by intravascular ultrasound: clinical, angiographic, and morphologic features and post-percutaneous coronary intervention complications in patients with acute coronary syndromes Usefulness of intravascular ultrasound guidance in percutaneous coronary intervention with second-generation drug-eluting stents for chronic total occlusions (from the Multicenter Korean-Chronic Total Occlusion Registry) American College of Cardiology Clinical Expert Consensus Document on Standards for Acquisition, Measurement and Reporting of Intravascular Ultrasound Studies (IVUS). A report of the American College of Cardiology Task Force on Clinical Expert Consensus Documents Intravascular ultrasound-guided implantation of drug-eluting stents to improve outcome: a meta-analysis The role of integrated backscatter intravascular ultrasound in characterizing bare metal and drug-eluting stent restenotic neointima as compared to optical coherence tomography Intravascular ultrasound findings of early stent thrombosis after primary percutaneous intervention in acute myocardial infarction: a Harmonizing Outcomes with Revascularization and Stents in Acute Myocardial Infarction (HORIZONS-AMI) substudy Outcomes with intravascular ultrasound-guided stent implantation: a meta-analysis of randomized trials in the era of drug-eluting stents First-in-man evaluation of intravascular optical frequency domain imaging (OFDI) of Terumo: a comparison with intravascular ultrasound and quantitative coronary angiography Prospective application of pre-defined intravascular ultrasound criteria for assessment of intermediate left main coronary artery lesions results from the multicenter LITRO study

Original Research2014 Mar 20;9(11):1301-8.

JOURNAL:EuroIntervention. Article Link

Mechanical complications of everolimus-eluting stents associated with adverse events: an intravascular ultrasound study

Inaba S, Mintz GS, Yun KH et al. Keywords: intravascular ultrasound; everolimus-eluting stent; stent fracture

ABSTRACT


AIMS - Mechanical complications contribute to bare metal and first-generation drug-eluting stent (DES) failure. However, the importance of the mechanical complications of second-generation DES remains unclear. We report mechanical complications associated with everolimus-eluting stent (EES) failures.


METHODS AND RESULTS - We retrospectively analysed 177 consecutive EES-treated lesions in 136 patients who underwent intravascular ultrasound (IVUS) at follow-up. Mechanical complications were identified in 17 patients (five stable angina, 10 unstable angina, two non-ST-elevation myocardial infarction [NSTEMI] without angiographic thrombus). Fifteen (88.2%) were treated with repeat revascularisation. By IVUS, there were 16 focal (94.1%) and one diffuse (5.9%) in-stent restenoses. Complete stent fracture with separation was seen in only one, partial stent fracture with separation was seen in three, and in 13 there was longitudinal deformation (n=2) or stent strut fracture (n=11) with overlapping of the proximal and distal stent fragments. In 13 EES with evidence of overlapping in the setting of either fracture or deformation, there was a 35.5±12.2% smaller stent area compared to the adjacent proximal and distal stent fragments, and >50% neointimal hyperplasia in 12 (92.3%).

CONCLUSIONS - We found EES mechanical complications, often followed by longitudinal deformation or fracture leading to excessive neointimal hyperplasia, in-stent restenosis, and repeat revascularisation.