CBS 2019
CBSMD教育中心
English

IVUS Guidance

科研文章

荐读文献

Angiographic and clinical comparisons of intravascular ultrasound- versus angiography-guided drug-eluting stent implantation for patients with chronic total occlusion lesions: two-year results from a randomised AIR-CTO study Three-Year Outcomes of the ULTIMATE Trial Comparing Intravascular Ultrasound Versus Angiography-Guided Drug-Eluting Stent Implantation Clinical impact of PCSK9 inhibitor on stabilization and regression of lipid-rich coronary plaques: a near-infrared spectroscopy study Role of Proximal Optimization Technique Guided by Intravascular Ultrasound on Stent Expansion, Stent Symmetry Index, and Side-Branch Hemodynamics in Patients With Coronary Bifurcation Lesions Role of intravascular ultrasound in patients with acute myocardial infarction undergoing percutaneous coronary intervention Subclinical Atherosclerosis Burden by 3D Ultrasound in Mid-Life: The PESA Study Tissue characterisation of atherosclerotic plaque in the left main: an in vivo intravascular ultrasound radiofrequency data analysis Impact of intravascular ultrasound on the long-term clinical outcomes in the treatment of coronary ostial lesions Utility of intravascular ultrasound guidance in patients undergoing percutaneous coronary intervention for type C lesions A Randomized Study of Distal Filter Protection Versus Conventional Treatment During Percutaneous Coronary Intervention in Patients With Attenuated Plaque Identified by Intravascular Ultrasound

Original Research2011 Jun;4(3):239-47.

JOURNAL:Circ Cardiovasc Interv. Article Link

Intravascular ultrasound findings of early stent thrombosis after primary percutaneous intervention in acute myocardial infarction: a Harmonizing Outcomes with Revascularization and Stents in Acute Myocardial Infarction (HORIZONS-AMI) substudy

Choi SY, Witzenbichler B, Maehara A et al. Keywords: stents; thrombosis; ultrasonics;myocardial infarction

ABSTRACT


BACKGROUND - Small stent area and residual inflow/outflow disease have been reported as the strongest intravascular ultrasound (IVUS) predictors of early stent thrombosis (ST) in patients with stable angina. IVUS predictors of early ST in patients with acute myocardial infarction have not been studied.


METHODS AND RESULTS - In the Harmonizing Outcomes with Revascularization and Stents in Acute Myocardial Infarction (HORIZONS-AMI) study, a formal substudy included poststent and 13-month follow-up IVUS at 36 centers. Twelve patients with baseline IVUS who had definite/probable early ST ≤30 days after enrollment were compared with 389 patients without early ST. Significant residual stenosis was a lumen area <4.0 mm(2) with ≥70% plaque burden ≤10 mm from each stent edge. Significant edge dissection was more than medial dissection with lumen area <4 mm(2) or dissection angle ≥60°. Randomization to bivalirudin (P=0.29) or paclitaxel-eluting stent (P=0.74) was not related to early ST. Minimum lumen area was smaller in patients with versus without early ST (4.4 mm(2) [3.6, 6.9] versus 6.7 mm(2) [5.3, 8.0], respectively, P=0.014). Minimum lumen area <5 mm(2), significant residual stenosis, significant stent edge dissection, and significant tissue (plaque/thrombus) protrusion (more than the median that narrowed the lumen to <4 mm(2)) were more prevalent in patients with early ST, but significant acute malapposition (more than the median) was not. Overall, 100% of patients with early ST had at least 1 of these significant features: minimum lumen area <5 mm(2), edge dissection, residual stenosis, or tissue protrusion versus 23% in patients without early ST (P<0.01).

CONCLUSIONS - Smaller final lumen area and inflow/outflow disease (residual stenosis or dissection) but not acute malapposition were related to early ST after acute myocardial infarction intervention.

CLINICAL TRIAL REGISTRATION - URL: http://www.clinicaltrials.gov. Unique identifier: NCT00433966.