CBS 2019
CBSMD教育中心
English

IVUS Guidance

科研文章

荐读文献

Clinical Outcomes Following Intravascular Imaging-Guided Versus Coronary Angiography-Guided Percutaneous Coronary Intervention With Stent Implantation: A Systematic Review and Bayesian Network Meta-Analysis of 31 Studies and 17,882 Patients Intravascular ultrasound guidance of percutaneous coronary intervention in ostial chronic total occlusions: a description of the technique and procedural results Clinical impact of intravascular ultrasound-guided chronic total occlusion intervention with zotarolimus-eluting versus biolimus-eluting stent implantation: randomized study Long-term health outcome and mortality evaluation after invasive coronary treatment using drug eluting stents with or without the IVUS guidance. Randomized control trial. HOME DES IVUS Intravascular ultrasound-guided unprotected left main coronary artery stenting in the elderly Effect of Intravascular Ultrasound-Guided vs Angiography-Guided Everolimus-Eluting Stent Implantation: The IVUS-XPL Randomized Clinical Trial Angiographic and clinical comparisons of intravascular ultrasound- versus angiography-guided drug-eluting stent implantation for patients with chronic total occlusion lesions: two-year results from a randomised AIR-CTO study IVUS Guidance for Coronary Revascularization: When to Start, When to Stop? A prospective, randomized trial of intravascular-ultrasound guided compared to angiography guided stent implantation in complex coronary lesions: the AVIO trial Subclinical Atherosclerosis Burden by 3D Ultrasound in Mid-Life: The PESA Study

Clinical Trial2011 Apr;6(9):1037-45.

JOURNAL:EuroIntervention. Article Link

First-in-man evaluation of intravascular optical frequency domain imaging (OFDI) of Terumo: a comparison with intravascular ultrasound and quantitative coronary angiography

Okamura T, Onuma Y, Garcia-Garcia HM et al. Keywords: IVUS; stents; OCT; coronary imaging; optical frequency domain imaging

ABSTRACT


AIMS The objective of this study is to evaluate the feasibility and safety of imaging human coronary arteries in vivo by optical frequency domain imaging (OFDI) in comparison to intravascular ultrasound (IVUS). OFDI has been recently developed to overcome the limitations of conventional time-domain optical coherence tomography (OCT), namely the need for proximal balloon occlusion. The Terumo-OFDI system is capable of acquiring images with high-speed automated pullback (up to 40 mm/sec) and requires only a short injection (3-4 sec) of small amount of x-ray contrast (9-16 ml).


METHODS AND RESULTS - Nineteen patients who underwent stent implantation were enrolled. IVUS/OFDI were performed before and after stenting. The incidences of any adverse event and angiographic adverse findings were recorded. Lumen area (LA) was measured by IVUS and OFDI at 1 mm intervals in the stented segments (n=19) as well as in the proximal, distal, and to-be-stented segments (n=40). In addition, lumen area in the stented segment was also measured by edge (E-) and video-densitometric (VD-) quantitative coronary angiography (QCA). The OFDI images were obtained without any adverse event related to imaging procedures. Post stenting (n=19), minimal LA (MLA) measured by OFDI (5.84 ± 1.89 mm2) was larger than that of E-QCA (4.16 ± 1.46 mm2, p<0.001) and VD-QCA (4.92 ± 1.55 mm2, p<0.05). It was smaller than IVUS-MLA (6.26 ± 2.01 mm2, N.S.) but the correlation between the two measurements was highly significant (R2=0.82, p<0.001).

CONCLUSIONS - The OFDI imaging is feasible both before and after stenting and has a promising safety profile. The OFDI provided clear high resolution images and robust lumen measurements.