Catastrophic catheter-induced coronary artery vasospasm successfully rescued using intravascular ultrasound imaging guidance Coronary plaque redistribution after stent implantation is determined by lipid composition: A NIRS-IVUS analysis Effect of Intravascular Ultrasound-Guided vs Angiography-Guided Everolimus-Eluting Stent Implantation: The IVUS-XPL Randomized Clinical Trial Comprehensive intravascular ultrasound assessment of stent area and its impact on restenosis and adverse cardiac events in 403 patients with unprotected left main disease Comparison of paclitaxel-eluting stents (Taxus) and everolimus-eluting stents (Xience) in left main coronary artery disease with 3 years follow-up (from the ESTROFA-LM registry) Meta-analysis of outcomes after intravascular ultrasound-guided versus angiography-guided drug-eluting stent implantation in 26,503 patients enrolled in three randomized trials and 14 observational studies Intravascular Ultrasound-Guided Versus Angiography-Guided Implantation of Drug-Eluting Stent in All-Comers: The ULTIMATE trial Increased glycated albumin and decreased esRAGE levels in serum are related to negative coronary artery remodeling in patients with type 2 diabetes: an Intravascular ultrasound study The Role of Vascular Imaging in Guiding Routine Percutaneous Coronary Interventions: A Meta-Analysis of Bare Metal Stent and Drug-Eluting Stent Trials Coronary plaque redistribution after stent implantation is determined by lipid composition: A NIRS-IVUS analysis

Clinical Trial2013 Mar;6(3):228-36.

JOURNAL:JACC Cardiovasc Interv. Article Link

Optical coherence tomography versus intravascular ultrasound to evaluate coronary artery disease and percutaneous coronary intervention

Bezerra HG, Attizzani GF, Sirbu V et al. Keywords: IVUS; OCT; PCI

ABSTRACT


OBJECTIVESWe compared intravascular ultrasound (IVUS) and 2 different generations of optical coherence tomography (OCT)-time-domain OCT (TD-OCT) and frequency-domain OCT (FD-OCT)-for the assessment of coronary disease and percutaneous coronary intervention (PCI) using stents.


BACKGROUND - OCT is a promising light-based intravascular imaging modality with higher resolution than IVUS. However, the paucity of data on OCT image quantification has limited its application in clinical practice.

METHODS - A total of 227 matched OCT and IVUS pull backs were studied. One hundred FD-OCT and IVUS pull backs in nonstented (n = 56) and stented (n = 44) vessels were compared. Additionally, 127 matched TD-OCT and IVUS images were compared in stented vessels.

RESULTS - FD-OCT depicted more severe native coronary disease than IVUS; minimal lumen area (MLA) was 2.33 ± 1.56 mm(2) versus 3.32 ± 1.92 mm(2), respectively (p < 0.001). Reference vessel dimensions were equivalent between FD-OCT and IVUS in both native and stented coronaries, but TD-OCT detected smaller reference lumen size compared with IVUS. Immediately post-PCI, in-stent MLAs were similar between FD-OCT and IVUS, but at follow-up, both FD-OCT and TD-OCT detected smaller MLAs than did IVUS, likely due to better detection of neointimal hyperplasia (NIH). Post-PCI malapposition and tissue prolapse were more frequently identified by FD-OCT.

CONCLUSIONS - FD-OCT generates similar reference lumen dimensions but higher degrees of disease severity and NIH, as well as better detection of malapposition and tissue prolapse compared with IVUS. First-generation TD-OCT was associated with smaller reference vessel dimensions compared with IVUS.

Copyright © 2013 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.