Higher neutrophil-to-lymphocyte ratio (NLR) increases the risk of suboptimal platelet inhibition and major cardiovascular ischemic events among ACS patients receiving dual antiplatelet therapy with ticagrelor Dual Antiplatelet TherapyIs It Time to Cut the Cord With Aspirin? Safety and efficacy of the bioabsorbable polymer everolimus-eluting stent versus durable polymer drug-eluting stents in high-risk patients undergoing PCI: TWILIGHT-SYNERGY Long-term dual antiplatelet-induced intestinal injury resulting in translocation of intestinal bacteria into blood circulation increased the incidence of adverse events after PCI in patients with coronary artery disease A risk score to predict postdischarge bleeding among acute coronary syndrome patients undergoing percutaneous coronary intervention: BRIC-ACS study Global Approach to High Bleeding Risk Patients With Polymer-Free Drug-Coated Coronary Stents: The LF II Study Sex-Based Outcomes in Patients With a High Bleeding Risk After Percutaneous Coronary Intervention and 1-Month Dual Antiplatelet Therapy: A Secondary Analysis of the LEADERS FREE Randomized Clinical Trial One-year outcome of a prospective trial stopping dual antiplatelet therapy at 3 months after everolimus-eluting cobalt-chromium stent implantation: ShortT and OPtimal duration of Dual AntiPlatelet Therapy after everolimus-eluting cobalt-chromium stent (STOPDAPT) trial Conceptual Framework for Addressing Residual Atherosclerotic Cardiovascular Disease Risk in the Era of Precision Medicine Ticagrelor With or Without Aspirin After Complex PCI

Clinical Trial2013 Oct;6(10):1095-1104.

JOURNAL:JACC Cardiovasc Imaging. Article Link

OCT compared with IVUS in a coronary lesion assessment: the OPUS-CLASS study

Kubo T, Akasaka T, Shite J et al. Keywords: FD; IVUS; MLA; MLD; OCT; PCI; QCA; TD; coronary angiography; frequency domain; intravascular ultrasound; minimum lumen area; minimum lumen diameter; optical coherence tomography; percutaneous coronary intervention; quantitative coronary angiography; time-domain

ABSTRACT


OBJECTIVESThe aim of this study was to investigate the reliability of frequency domain optical coherence tomography (FD-OCT) for coronary measurements compared with quantitative coronary angiography (QCA) and intravascular ultrasound (IVUS).


BACKGROUND - Accurate luminal measurement is expected in FD-OCT because this technology offers high resolution and excellent contrast between lumen and vessel wall.

METHODS - In 5 medical centers, 100 patients with coronary artery disease were prospectively studied by using angiography, FD-OCT, and IVUS. In addition, 5 phantom models of known lumen dimensions (lumen diameter 3.08 mm; lumen area 7.45 mm(2)) were examined using FD-OCT and IVUS. Quantitative image analyses of the coronary arteries and phantom models were performed by an independent core laboratory.

RESULTS - In the clinical study, the mean minimum lumen diameter measured by QCA was significantly smaller than that measured by FD-OCT (1.81 ± 0.72 mm vs. 1.91 ± 0.69 mm; p < 0.001) and the minimum lumen diameter measured by IVUS was significantly greater than that measured by FD-OCT (2.09 ± 0.60 mm vs. 1.91 ± 0.69 mm; p < 0.001). The minimum lumen area measured by IVUS was significantly greater than that by FD-OCT (3.68 ± 2.06 mm(2) vs. 3.27 ± 2.22 mm(2); p < 0.001), although a significant correlation was observed between the 2 imaging techniques (r = 0.95, p < 0.001; mean difference 0.41 mm(2)). Both FD-OCT and IVUS exhibited good interobserver reproducibility, but the root-mean-squared deviation between measurements was approximately twice as high for the IVUS measurements compared with the FD-OCT measurements (0.32 mm(2) vs. 0.16 mm(2)). In a phantom model, the mean lumen area according to FD-OCT was equal to the actual lumen area of the phantom model, with low SD; IVUS overestimated the lumen area and was less reproducible than FD-OCT (8.03 ± 0.58 mm(2) vs. 7.45 ± 0.17 mm(2); p < 0.001).

CONCLUSIONS - The results of this prospective multicenter study demonstrate that FD-OCT provides accurate and reproducible quantitative measurements of coronary dimensions in the clinical setting.

Copyright © 2013 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.