Right ventricular function and outcome in patients undergoing transcatheter aortic valve replacement Management of Asymptomatic Severe Aortic Stenosis: Evolving Concepts in Timing of Valve Replacement Single Versus Dual Antiplatelet Therapy Following TAVR: A Systematic Review and Meta-Analysis of Randomized Controlled Trials 2-Year Outcomes After Transcatheter Versus Surgical Aortic Valve Replacement in Low-Risk Patients Contemporary Presentation and Management of Valvular Heart Disease: The EURObservational Research Programme Valvular Heart Disease II Survey Increased Risk of Valvular Heart Disease in Systemic Sclerosis: An Underrecognized Cardiac Complication Impact of Severe Sarcopenia on Rehospitalization and Survival One Year After a TAVR Procedure in Patients Aged 75 and Older Anticoagulation with or without Clopidogrel after Transcatheter Aortic-Valve Implantation Chimney technique in a TAVR-in-TAVR procedure with high risk of left main artery ostium occlusion Delirium After TAVR: Crosspassing the Limit of Resilience

Original ResearchOctober 7, 2020

JOURNAL:JAMA Cardiol. Article Link

Transcatheter Aortic Valve Replacement in Low-risk Patients With Bicuspid Aortic Valve Stenosis

JK Forrest, B Ramlawi, GM Deeb et al.

ABSTRACT

IMPORTANCE - The outcomes of transcatheter aortic valve replacement (TAVR) in low-risk patients with bicuspid aortic valve stenosis have not been studied in a large scale, multicentered, prospective fashion.


OBJECTIVE - To evaluate the procedural safety, efficacy, and 30-day outcomes of TAVR in patients with bicuspid aortic stenosis at low surgical risk.


DESIGN, SETTING, AND PARTICIPANTS - The Low Risk Bicuspid Study is a prospective, single-arm trial study with inclusion/exclusion criteria developed from the Evolut Low Risk Randomized Trial. Follow-up is planned for 10 years. Patients underwent TAVR at 25 centers in the United States who were also participating in the Evolut Low Risk Randomized Trial from December 2018 to October 2019. Eligible patients had severe bicuspid aortic valve stenosis and met American Heart Association/American College of Cardiology guideline indications for aortic valve replacement.

 

INTERVENTIONS - Patients underwent attempted implant of an Evolut or Evolut PRO transcatheter aortic valve, with valve size based on annular measurements.

 

MAIN OUTCOMES AND MEASURES - The prespecified primary end point was the incidence of all-cause mortality or disabling stroke at 30 days. The prespecified primary efficacy end point was device success defined as the absence of procedural mortality, the correct position of 1 bioprosthetic heart valve in the proper anatomical location, and the absence of more than mild aortic regurgitation postprocedure.

 

RESULTS - A total of 150 patients underwent an attempted implant. Baseline characteristics include mean age of 70.3 (5.5) years, 48.0% female (n = 72), and a mean Society of Thoracic Surgeons score of 1.4 (0.6%). Most patients (136; 90.7%) had Sievers type I valve morphology. The incidence of all-cause mortality or disabling stroke was 1.3% (95% CI, 0.3%-5.3%) at 30 days. The device success rate was 95.3% (95% CI, 90.5%-98.1%). At 30 days, the mean (SD) AV gradient was 7.6 (3.7) mm Hg and effective orifice area was 2.3 (0.7) cm2. A new permanent pacemaker was implanted in 22 patients (15.1%). No patients had greater than mild paravalvular leak.

 

CONCLUSIONS AND RELEVANCE - Transcatheter aortic valve replacement in lowsurgical risk patients with bicuspid aortic valve stenosis achieved favorable 30-day results, with low rates of death and stroke and high device success rate.

 