CBS 2019
CBSMD教育中心
English

Percutaneous LAA Occlusion

科研文章

荐读文献

Comparison of the Efficacy and Safety Outcomes of Edoxaban in 8040 Women Versus 13 065 Men With Atrial Fibrillation in the ENGAGE AF-TIMI 48 Trial Anticoagulation in Concomitant Chronic Kidney Disease and Atrial Fibrillation: JACC Review Topic of the Week Short-Term Oral Anticoagulation Versus Antiplatelet Therapy Following Transcatheter Left Atrial Appendage Closure Gut microbiota dysbiosis promotes age-related atrial fibrillation by lipopolysaccharide and glucose-induced activation of NLRP3-inflammasome Edoxaban-based versus vitamin K antagonist-based antithrombotic regimen after successful coronary stenting in patients with atrial fibrillation (ENTRUST-AF PCI): a randomised, open-label, phase 3b trial Early Rhythm-Control Therapy in Patients with Atrial Fibrillation Alcohol consumption, cardiac biomarkers, and risk of atrial fibrillation and adverse outcomes Left Atrial Appendage Occlusion during Cardiac Surgery to Prevent Stroke Frailty and Clinical Outcomes of Direct Oral Anticoagulants Versus Warfarin in Older Adults With Atrial Fibrillation: A Cohort Study Detection of Device-Related Thrombosis Following Left Atrial Appendage Occlusion A Comparison Between Cardiac Computed Tomography and Transesophageal Echocardiography​: A Comparison Between Cardiac Computed Tomography and Transesophageal Echocardiography

Original Research2020 Nov 4;S1936-8798(20)31468-0.

JOURNAL:JACC Cardiovasc Interv. Article Link

Procedural and Short-Term Results With the New Watchman FLX Left Atrial Appendage Occlusion Device

Ignacio Cruz-Gonzalez 1 , Kasper Korsholm 2 , Blanca Trejo-Velasco Keywords: Watchman; left atrial appendage occlusion

ABSTRACT

OBJECTIVES - This study sought to report early experience with the new-generation Watchman FLX device (Boston Scientific, Marlborough, Massachusetts).

BACKGROUND - The new-generation Watchman FLX features a reduced height, improved anchoring and fabric coverage, and a closed distal end. These design modifications aim to simplify implantation, allow full recapture and repositioning, and reduce peridevice leak and device-related thrombosis.

METHODS - A total of 165 patients undergoing left atrial appendage (LAA) occlusion (LAAO) with Watchman FLX were enrolled in a prospective, multicenter registry at 12 centers participating in the European limited market release program.

RESULTS - Mean age was 75.4 ± 8.9 years, and CHA2DS2-VASc score 4.4 ± 1.4. A total of 128 patients (77.6%) had a history of major bleeding, including previous intracranial hemorrhage in 55 cases (33.3%). LAA landing zone minimal and maximal mean diameters were 19.1 ± 3.6 mm and 22.3 ± 3.7 mm, and 24.2% of LAA were considered complex by dimensions. Technical success was achieved in all patients. Successful implantation at first attempt was achieved in 129 cases (78.2%), and a second device was required in 6 cases (3.6%). Procedure-related complications occurred in 3 patients (1.8%): 2 access-related (1.2%) and 1 pericardial effusion (0.6%). No peri-procedural strokes, deaths, or device embolizations occurred. Forty-nine patients (29.7%) were discharged with single antiplatelet therapy, 105 (63.6%) on dual antiplatelet, and 11 (6.7%) on anticoagulation. Imaging follow-up displayed just 1 peridevice leak 5 mm and 7 cases of device-related thrombosis (4.7%). During a median follow-up of 55 days (interquartile range: 45 to 148 days), there were 6 hemorrhagic complications (4.8%), 1 patient (0.8%) had an ischemic stroke, and 1 (0.8%) died. No late device embolizations occurred.

CONCLUSIONS - LAAO with the Watchman FLX is safe and effective in a wide range of LAA morphologies, with a low procedural complication rate, high degree of LAA sealing, and favorable short-term efficacy.