Second vs. First generation drug eluting stents in multiple vessel disease and left main stenosis: Two-year follow-up of the observational, prospective, controlled, and multicenter ERACI IV registry Clinical Outcome After DK Crush Versus Culotte Stenting of Distal Left Main Bifurcation Lesions: The 3-Year Follow-Up Results of the DKCRUSH-III Study Long-term results after PCI of unprotected distal left main coronary artery stenosis: the Bifurcations Bad Krozingen (BBK)-Left Main Registry Ten-Year All-Cause Death According to Completeness of Revascularization in Patients With Three-Vessel Disease or Left Main Coronary Artery Disease: Insights From the SYNTAX Extended Survival Study Impact of Staging Percutaneous Coronary Intervention in Left Main Artery Disease: Insights From the EXCEL Trial Why NOBLE and EXCEL Are Consistent With Each Other and With Previous Trials Complex PCI procedures: challenges for the interventional cardiologist Comparison of Outcomes of Percutaneous Coronary Intervention on Native Coronary Arteries Versus on Saphenous Venous Aorta Coronary Conduits in Patients With Low Left Ventricular Ejection Fraction and Impella Device Implantation Achieved or Attempted (from the PROTECT II Randomized Trial and the cVAD Registry)

Clinical TrialAugust 2018. [Epub ahead]

JOURNAL:JACC Cardiovasc Interv. Article Link

A Randomized Study of Distal Filter Protection Versus Conventional Treatment During Percutaneous Coronary Intervention in Patients With Attenuated Plaque Identified by Intravascular Ultrasound

K Hibi, K Kozuma, S Sonoda et al. Keywords: acute coronary syndrome(s); distal embolism; distal protection device; intravascular ultrasound

ABSTRACT


OBJECTIVES - The aim of this study was to evaluate the utility of distal protection during percutaneous coronary intervention (PCI) in patients with acute coronary syndromes at high risk for distal embolization.


BACKGROUND - The results of previous clinical trials indicated that the routine use of distal protection in patients with ST-segment elevation myocardial infarction did not improve clinical outcomes. However, selective use of distal protection by means of a filter-based distal protection system has not been evaluated.

METHODS - Two hundred patients with acute coronary syndromes who had native coronary artery lesions and attenuated plaque with longitudinal length ≥5 mm on pre-PCI intravascular ultrasound were randomly assigned to undergo PCI with distal protection or conventional treatment.

RESULTS - The primary endpoint (no-reflow phenomenon) occurred in 26 patients (26.5%) in the distal protection group and 39 patients (41.7%) in the conventional treatment group (p = 0.026), and the corrected TIMI (Thrombolysis In Myocardial Infarction) frame count after revascularization was significantly lower in the distal protection group (23 vs. 30.5; p = 0.0003). The incidence of cardiac death, cardiac arrest, cardiogenic shock after revascularization requiring defibrillation, cardiopulmonary resuscitation, or extracorporeal membrane oxygenation was significantly lower in the distal protection group than in the conventional treatment group (0% vs. 5.2%; p = 0.028).

CONCLUSIONS - The use of distal embolic protection applied with a filter device decreased the incidence of the no-reflow phenomenon and was associated with fewer serious adverse cardiac events after revascularization than conventional PCI in patients with acute coronary syndromes with attenuated plaque ≥5 mm in length. (Assessment of Distal Protection Device in Patients at High Risk for Distal Embolism in Acute Coronary Syndrome [ACS] [VAMPIRE3]; NCT01460966)