Optical coherence tomography is a kid on the block: I would choose intravascular ultrasound Intravascular ultrasound guidance of percutaneous coronary intervention in ostial chronic total occlusions: a description of the technique and procedural results IVUS Guidance Is Associated With Better Outcome in Patients Undergoing Unprotected Left Main Coronary Artery Stenting Compared With Angiography Guidance Alone Successful Treatment of Unprotected Left Main Coronary Bifurcation Lesion Using Minimum Contrast Volume with Intravascular Ultrasound Guidance Intravascular ultrasound guidance in drug-eluting stents implantation: a meta-analysis and trial sequential analysis of randomized controlled trials Intravascular ultrasound-guided percutaneous coronary intervention in left main coronary bifurcation lesions: a review Clinical Outcomes Following Intravascular Imaging-Guided Versus Coronary Angiography–Guided Percutaneous Coronary Intervention With Stent Implantation: A Systematic Review and Bayesian Network Meta-Analysis of 31 Studies and 17,882 Patients In Vivo Calcium Detection by Comparing Optical Coherence Tomography, Intravascular Ultrasound, and Angiography

Clinical Trial2018 Jan 1;19(1):59-66.

JOURNAL:Eur Heart J Cardiovasc Imaging. Article Link

Fate of post-procedural malapposition of everolimus-eluting polymeric bioresorbable scaffold and everolimus-eluting cobalt chromium metallic stent in human coronary arteries: sequential assessment with optical coherence tomography in ABSORB Japan trial

Sotomi Y, Onuma Y, Dijkstra J et al. Keywords: bioresorbable scaffold ; metallic stent ; optical coherence tomography ; randomized controlled trial; strut malapposition

ABSTRACT


AIMS - The natural course of post-procedural incomplete strut apposition (ISA) after the implantation of bioresorbable scaffolds (BVS) remains unknown. The purpose of the present study was to evaluate the fate of post-procedural ISA after everolimus-eluting Absorb BVS in comparison with the second-generation everolimus-eluting cobalt chromium stent (CoCr-EES).


METHODS AND RESULTS - Fate of post-procedural ISA was evaluated by serial optical coherence tomography(OCT) in the ABSORB Japan randomized trial [OCT-1 subgroup: 110 paired lesions of post-procedure and 2-year follow-up (BVS 73 lesions vs. CoCr-EES 37 lesions)] with respect to ISA distance. Post-procedure ISA struts were categorized into either 'resolved' or 'persistent' by matched OCT imaging at 2-year follow-up. Post-procedure %malapposed strut and ISA area were smaller in BVS than in CoCr-EES (%malapposed strut: 4.8 ± 6.9% vs. 9.9 ± 9.8%, P = 0.002; ISA area 0.10 ± 0.18 mm2 vs. 0.23 ± 0.26 mm2, P = 0.003). At 2-year follow-up, the difference diminished, and majority of the ISA struts spontaneously resolved in both arms (%malapposed strut: 0.10 ± 0.46% vs. 0.24 ± 0.65%, P = 0.183). Receiver operating characteristic curve analysis (BVS 661 struts vs. CoCr-EES 807 struts) demonstrated that the best cut-off value of endoluminal ISA distance post-procedure for predicting persistent-ISA at 2-year follow-up was 396 µm for BVS (sensitivity 0.875; specificity 0.851) and 359 µm for CoCr-EES (sensitivity 0.778; specificity 0.881).


CONCLUSION - BVS as compared with CoCr-EES allows larger ISA distance at post-procedure, although we should make every effort to minimize post-procedure ISA. The reported cut-off value of OCT-estimated ISA distance at post-stenting for predicting persistent-ISA would be helpful to optimize PCI with BVS and CoCr-EES.