CBS 2019
CBSMD教育中心
中 文

血管内超声指导

Abstract

Recommended Article

Serial intravascular ultrasound assessment of very late stent thrombosis after sirolimus-eluting stent placement Coronary artery imaging with intravascular high-frequency ultrasound Usefulness of minimum stent cross sectional area as a predictor of angiographic restenosis after primary percutaneous coronary intervention in acute myocardial infarction (from the HORIZONS-AMI Trial IVUS substudy) Randomized comparison of clinical outcomes between intravascular ultrasound and angiography-guided drug-eluting stent implantation for long coronary artery stenoses Clinical impact of PCSK9 inhibitor on stabilization and regression of lipid-rich coronary plaques: a near-infrared spectroscopy study American College of Cardiology Clinical Expert Consensus Document on Standards for Acquisition, Measurement and Reporting of Intravascular Ultrasound Studies (IVUS). A report of the American College of Cardiology Task Force on Clinical Expert Consensus Documents Intravascular Ultrasound and Angioscopy Assessment of Coronary Plaque Components in Chronic Totally Occluded Lesions Impact of Positive and Negative Lesion Site Remodeling on Clinical Outcomes : Insights From PROSPECT

Original Research25 May 2021

JOURNAL:Eurointervention. Article Link

Incidence and Standardized Definitions of Mitral Valve Leaflet Adverse Events After Transcatheter Mitral Valve Repair: the EXPAND Study

FM Asch, SH Little, GB Mackensen et al. Keywords: mitral valve leaflet adverse events; LAE; definition; MitraClip;

ABSTRACT

BACKGROUND - An independent panel of experts reviewed all investigator-reported cases of mitral valve leaflet adverse events (LAE) after MitraClipTM NTR/XTR in the EXPAND Study.


AIMS - We aimed to report the findings of the expert panel and standardize definitions for LAE.


METHODS - Standard definitions for different types of LAE were formulated and events adjudicated after detailed review by the expert panel.


RESULTS - Enrolling centers reported LAE in 35 cases, 11 leaflet injuries (9 tear, 2 perforation) and 24 single leaflet device attachment (SLDA). The panel confirmed LAE in 20 cases (2.0% incidence), 18 patients had SLDA and 4 had leaflet injury (2 cases had both SLDA and injury). Leaflet injury occurred during device implant and resulted in surgical valve replacement or death. SLDA-alone events were identified during implant (n=2), pre-discharge (7) or at 30 days of follow-up (7) and were resolved (£ 2+ residual MR) with additional clips in 75% of cases.


CONCLUSIONS - Mitral valve repair with MitraClipTM NTR/XTR is safe. The rate of LAE is lower than previously reported using older generation devices. The proposed definitions and findings will help differentiate leaflet injury from inadequate leaflet insertion and SLDA, and provide guidance to consistently diagnose LAE post MitraClipTM.