Serial intravascular ultrasound assessment of very late stent thrombosis after sirolimus-eluting stent placement Usefulness of minimum stent cross sectional area as a predictor of angiographic restenosis after primary percutaneous coronary intervention in acute myocardial infarction (from the HORIZONS-AMI Trial IVUS substudy) Randomized comparison of clinical outcomes between intravascular ultrasound and angiography-guided drug-eluting stent implantation for long coronary artery stenoses Coronary artery imaging with intravascular high-frequency ultrasound Clinical impact of PCSK9 inhibitor on stabilization and regression of lipid-rich coronary plaques: a near-infrared spectroscopy study Intravascular Ultrasound and Angioscopy Assessment of Coronary Plaque Components in Chronic Totally Occluded Lesions American College of Cardiology Clinical Expert Consensus Document on Standards for Acquisition, Measurement and Reporting of Intravascular Ultrasound Studies (IVUS). A report of the American College of Cardiology Task Force on Clinical Expert Consensus Documents Novel predictors of late lumen enlargement in distal reference segments after successful recanalization of coronary chronic total occlusion

Clinical Trial2013 Mar;6(3):228-36.

JOURNAL:JACC Cardiovasc Interv. Article Link

Optical coherence tomography versus intravascular ultrasound to evaluate coronary artery disease and percutaneous coronary intervention

Bezerra HG, Attizzani GF, Sirbu V et al. Keywords: IVUS; OCT; PCI

ABSTRACT


OBJECTIVESWe compared intravascular ultrasound (IVUS) and 2 different generations of optical coherence tomography (OCT)-time-domain OCT (TD-OCT) and frequency-domain OCT (FD-OCT)-for the assessment of coronary disease and percutaneous coronary intervention (PCI) using stents.


BACKGROUND - OCT is a promising light-based intravascular imaging modality with higher resolution than IVUS. However, the paucity of data on OCT image quantification has limited its application in clinical practice.

METHODS - A total of 227 matched OCT and IVUS pull backs were studied. One hundred FD-OCT and IVUS pull backs in nonstented (n = 56) and stented (n = 44) vessels were compared. Additionally, 127 matched TD-OCT and IVUS images were compared in stented vessels.

RESULTS - FD-OCT depicted more severe native coronary disease than IVUS; minimal lumen area (MLA) was 2.33 ± 1.56 mm(2) versus 3.32 ± 1.92 mm(2), respectively (p < 0.001). Reference vessel dimensions were equivalent between FD-OCT and IVUS in both native and stented coronaries, but TD-OCT detected smaller reference lumen size compared with IVUS. Immediately post-PCI, in-stent MLAs were similar between FD-OCT and IVUS, but at follow-up, both FD-OCT and TD-OCT detected smaller MLAs than did IVUS, likely due to better detection of neointimal hyperplasia (NIH). Post-PCI malapposition and tissue prolapse were more frequently identified by FD-OCT.

CONCLUSIONS - FD-OCT generates similar reference lumen dimensions but higher degrees of disease severity and NIH, as well as better detection of malapposition and tissue prolapse compared with IVUS. First-generation TD-OCT was associated with smaller reference vessel dimensions compared with IVUS.