CBS 2019
CBSMD教育中心
中 文

急性冠脉综合征

Abstract

Recommended Article

Association between Coronary Collaterals and Myocardial Viability in Patients with a Chronic Total Occlusion Multivessel PCI Guided by FFR or Angiography for Myocardial Infarction High-Sensitivity Troponin and The Application of Risk Stratification Thresholds in Patients with Suspected Acute Coronary Syndrome Improved outcomes in patients with ST-elevation myocardial infarction during the last 20 years are related to implementation of evidence-based treatments: experiences from the SWEDEHEART registry 1995-2014 Cardiac Troponin Composition Characterization after Non ST-Elevation Myocardial Infarction: Relation with Culprit Artery, Ischemic Time Window, and Severity of Injury Effect of a Restrictive vs Liberal Blood Transfusion Strategy on Major Cardiovascular Events Among Patients With Acute Myocardial Infarction and Anemia: The REALITY Randomized Clinical Trial Clinical and Angiographic Features of Patients With Out-of-Hospital Cardiac Arrest and Acute Myocardial Infarction Revascularization Strategies in STEMI with Multivessel Disease: Deciding on Culprit Versus Complete-Ad Hoc or Staged

Clinical Trial2018 Aug 16.

JOURNAL:Cardiovasc Revasc Med. Article Link

A randomised trial comparing two stent sizing strategies in coronary bifurcation treatment with bioresorbable vascular scaffolds - The Absorb Bifurcation Coronary (ABC) trial

Rampat R, Mayo T, Hildick-Smith D et al. Keywords: Absorb; BVS; Bioresorbable vascular scaffold; Coronary bifurcation

ABSTRACT


BACKGROUNDLimited information is available on the use of Bioresorbable Vascular Scaffold (BVS) in bifurcations involving significant side branches. When treating bifurcation disease with metal stents, the recommendation is to choose a stent diameter based on the distal main vessel diameter. Whether this sizing strategy is applicable to BVS is currently unknown.


METHODS - We randomised 37 patients undergoing elective PCI for 'false' bifurcation disease (Medina 0,1,0; 1,0,0; 1,1,0) to receive BVS based either on proximal or distal reference diameters. Optical Frequency Domain Imaging (OFDI) measurements were performed pre BVS insertion to obtain proximal and distal reference diameters and post implantation. BVS size was chosen according to the proximal or distal reference diameter as per randomisation. Implantation was performed using the PSP technique tailored to bifurcation stenting. OFDI was repeated post implantation to confirm satisfactory expansion and apposition.

RESULTS - Baseline demographics between the two groups were similar. Patients were aged 62.8 ± 3.3 years; 76% were male. Mean side branch diameter was 2.24 ± 0.13 mm. TIMI III flow in the main vessel was achieved in all cases. Side branch occlusion occurred in 1 case (2.7%). In the distal-sizing arm, there was a greater incidence of significant malapposition (>300 μm) at the proximal end of the scaffold on OCT (2.3% versus 0.8%, p 0.023). The incidence of distal edge dissections was numerically greater in the proximal-sizing group but this was not statistically significant (31.3% vs 11.8%, p 0.17).

CONCLUSION - Both proximal and distal sizing strategies have similar procedural complication rates when using the ABSORB BVS to treat coronary bifurcations. However a proximal sizing strategy is associated with less malapposition and may be preferable.

Copyright © 2018 Elsevier Inc. All rights reserved.