CBS 2019
CBSMD教育中心
中 文

急性冠脉综合征

Abstract

Recommended Article

Impact of Off-Hours Versus On-Hours Primary Percutaneous Coronary Intervention on Myocardial Damage and Clinical Outcomes in ST-Segment Elevation Myocardial Infarction Canadian SCAD Cohort Study: Shedding Light on SCAD From a United Front Association of the PHACTR1/EDN1 Genetic Locus With Spontaneous Coronary Artery Dissection Effect of Smoking on Outcomes of Primary PCI in Patients With STEMI BMI, Infarct Size, and Clinical Outcomes Following Primary PCI Patient-Level Analysis From 6 Randomized Trials Decreased inspired oxygen stimulates de novo formation of coronary collaterals in adult heart Intraaortic Balloon Pump in Cardiogenic Shock Complicating Acute Myocardial Infarction: Long-Term 6-Year Outcome of the Randomized IABP-SHOCK II Trial Advances in Clinical Cardiology 2020: A Summary of Key Clinical Trials

Clinical Trial2018 Sep 27;379(13):1205-1215.

JOURNAL:N Engl J Med. Article Link

Wearable Cardioverter-Defibrillator after Myocardial Infarction

Olgin JE, Pletcher MJ, VEST Investigators et al. Keywords: wearable cardioverter–defibrillator; myocardial infarction; sudden death prevention

ABSTRACT


BACKGROUND - Despite the high rate of sudden death after myocardial infarction among patients with a low ejection fraction, implantable cardioverter-defibrillators are contraindicated until 40 to 90 days after myocardial infarction. Whether a wearable cardioverter-defibrillator would reduce the incidence of sudden death during this high-risk period is unclear.


METHODS - We randomly assigned (in a 2:1 ratio) patients with acute myocardial infarction and an ejection fraction of 35% or less to receive a wearable cardioverter-defibrillator plus guideline-directed therapy (the device group) or to receive only guideline-directed therapy (the control group). The primary outcome was the composite of sudden death or death from ventricular tachy arrhythmia at 90 days (arrhythmic death). Secondary outcomes included death from any cause and nonarrhythmic death.


RESULTS - Of 2302 participants, 1524 were randomly assigned to the device group and 778 to the control group. Participants in the device group wore the device for a median of 18.0 hours per day (interquartile range, 3.8 to 22.7). Arrhythmic death occurred in 1.6% of the participants in the device group and in 2.4% of those in the control group (relative risk, 0.67; 95% confidence interval [CI], 0.37 to 1.21; P=0.18). Death from any cause occurred in 3.1% of the participants in the device group and in 4.9% of those in the control group (relative risk, 0.64; 95% CI, 0.43 to 0.98; uncorrected P=0.04), and nonarrhythmic death in 1.4% and 2.2%, respectively (relative risk, 0.63; 95% CI, 0.33 to 1.19; uncorrected P=0.15). Of the 48 participants in the device group who died, 12 were wearing the device at the time of death. A total of 20 participants in the device group (1.3%) received an appropriate shock, and 9 (0.6%) received an inappropriate shock.


CONCLUSIONS - Among patients with a recent myocardial infarction and an ejection fraction of 35% or less, the wearable cardioverter-defibrillator did not lead to a significantly lower rate of the primary outcome of arrhythmic death than control. (Funded by the National Institutes of Health and Zoll Medical; VEST ClinicalTrials.gov number, NCT01446965 .).