BMI, Infarct Size, and Clinical Outcomes Following Primary PCI Patient-Level Analysis From 6 Randomized Trials Comparison of Inhospital Mortality and Frequency of Coronary Angiography on Weekend Versus Weekday Admissions in Patients With Non-ST-Segment Elevation Acute Myocardial Infarction Late Survival Benefit of Percutaneous Coronary Intervention Compared With Medical Therapy in Patients With Coronary Chronic Total Occlusion: A 10-Year Follow-Up Study Interval From Initiation of Prasugrel to Coronary Angiography in Patients With Non–ST-Segment Elevation Myocardial Infarction Mortality in STEMI patients without standard modifiable risk factors: a sex-disaggregated analysis of SWEDEHEART registry data Antithrombotic Therapy in Patients With Atrial Fibrillation and Acute Coronary Syndrome Ticagrelor or Prasugrel in Patients with Acute Coronary Syndromes Advances in Clinical Cardiology 2020: A Summary of Key Clinical Trials

Original ResearchSeptember 2019

JOURNAL:JACC Cardiovasc Interv. Article Link

Long-Term Durability of Transcatheter Heart Valves: Insights From Bench Testing to 25 Years

J Sathananthan, M Hensey, U Landes et al. Keywords: durability; transcatheter aortic valve replacement; transcatheter heart valve

ABSTRACT


OBJECTIVES - This study assessed the long-term durability of nominally deployed transcatheter heart valves (THV) to 1 billion cycles (equivalent to 25 years) and non-nominal (overexpansion, underexpansion, and elliptical) THV deployments to 200 million cycles (equivalent to 5 years) with accelerated wear testing.


METHODS - SAPIEN 3 THVs, sized 20, 23, 26, and 29 mm were assessed. Nominally deployed THVs underwent hydrodynamic performance and mechanical durability as assessed with accelerated wear testing to 1 billion cycles. Magna Ease surgical valves were used as comparators. Durability of non-nominal THV deployments was tested to 200 million cycles. Valves were tested to International Standards Organization 5840:2013 standard.


RESULTS - THV durability was excellent for both the nominal and non-nominal THV deployments to 1 billion and 200 million cycles, respectively. At 1 billion cycles the regurgitant fraction for the 20-, 23-, 26-, and 29-mm SAPIEN 3 was 0.92 ± 0.47%, 1.29 ± 0.04%, 1.73 ± 0.46%, and 2.47 ± 0.15%, respectively. There was also excellent durability in the comparator Magna Ease valves. The regurgitant fraction of non-nominal overexpanded (20 mm, 4.36 ± 0.53; 23 mm, 7.68 ± 1.39; 26 mm, 6.80 ± 1.17; 29 mm, 9.00 ± 0.37), underexpanded (20 mm, 3.06 ± 0.28; 23 mm, 4.46 ± 0.45; 26 mm, 7.72 ± 0.48; 29 mm, 8.65 ± 2.01), and elliptical (20 mm, 3.30 ± 0.38; 23 mm, 6.13 ± 0.94; 26 mm, 6.77 ± 1.22; 29 mm, 8.72 ± 0.24) THVs were excellent at 200 million cycles.


CONCLUSIONS - Nominal SAPIEN 3 THVs demonstrated excellent durability, to an equivalent of 25-years wear. THV durability was similar to the comparator surgical valves tested. Non-nominal (overexpansion, underexpansion, and elliptical) THV deployments also had excellent durability to an equivalent of 5-years wear.