CBS 2019
CBSMD教育中心
中 文

急性冠脉综合征

Abstract

Recommended Article

Impact of Chronic Total Coronary Occlusion Location on Long-term Survival After Percutaneous Coronary Intervention Revascularization Strategies in STEMI with Multivessel Disease: Deciding on Culprit Versus Complete-Ad Hoc or Staged Incidence and Outcomes of Acute Coronary Syndrome After Transcatheter Aortic Valve Replacement Effect of Smoking on Outcomes of Primary PCI in Patients With STEMI Hospital Readmission After Perioperative Acute Myocardial Infarction Associated With Noncardiac Surgery Decreased inspired oxygen stimulates de novo formation of coronary collaterals in adult heart 2021 ACC/AHA/SCAI Guideline for Coronary Artery Revascularization: A Report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines Impact of door-to-balloon time on long-term mortality in high- and low-risk patients with ST-elevation myocardial infarction

Clinical TrialVolume 11, Issue 3, February 2018

JOURNAL:JACC Cardiovasc Interv. Article Link

A Randomized Trial Comparing the NeoVas Sirolimus-Eluting Bioresorbable Scaffold and Metallic Everolimus-Eluting Stents

YL Han, B Xu, NeoVas Randomized Controlled Trial Investigators Keywords: bioresorbable scaffolds; drug-eluting stent(s); randomized controlled trial;

ABSTRACT


OBJECTIVES - The authors sought to evaluate the safety and effectiveness of the NeoVas bioresorbable scaffold (BRS) compared with metallic drug-eluting stents.



BACKGROUND -  BRS have the potential to improve very late outcomes compared with metallic drug-eluting stents, but some BRS have been associated with increased rates of device thrombosis before complete bioresorption. NeoVas is a new poly-L-lactic acid BRS that elutes sirolimus from a poly-D, L-lactide coating.



METHODS -  Eligible patients with a single de novo native coronary artery lesion with a reference vessel diameter 2.5 to 3.75 mm and a lesion length ≤20 mm were randomized 1:1 to NeoVas BRS versus cobalt-chromium everolimus-eluting stents (CoCr-EES). Angiographic follow-up was performed in all patients at 1 year. The primary endpoint was angiographic in-segment late loss (LL), and the major secondary endpoint was the rate of angina. Baseline and follow-up optical coherence tomography and fractional flow reserve were performed in a pre-specified subgroup of patients.



RESULTS -  The authors randomized 560 patients at 32 centers to treatment with NeoVas (n = 278) versus CoCr-EES (n = 282). One-year in-segment LL with NeoVas and CoCr-EES were 0.14 ± 0.36 mm versus 0.11 ± 0.34 mm (difference 0.03 mm; upper 1-sided 97.5% confidence interval 0.09 mm; pnoninferiority < 0.0001; psuperiority = 0.36). Clinical outcomes at 1 year were similar in the 2 groups, as were the rates of recurrent angina (27.9% vs. 32.1%; p = 0.26). Optical coherence tomography at 1 year demonstrated a higher proportion of covered struts (98.7% vs. 96.2%; p < 0.001), less strut malapposition (0% vs. 0.6%; p <0.001), and a smaller minimal lumen area (4.71 ± 1.64 vs. 6.00 ± 2.15 mm2; p < 0.001) with NeoVas compared with CoCr-EES respectively, with nonsignificant differences in fractional flow reserve (0.89 ± 0.08 vs. 0.91 ± 0.06; p = 0.07).



CONCLUSIONS - The NeoVas BRS was noninferior to CoCr-EES for the primary endpoint of 1-year angiographic in-segment LL, and resulted in comparable 1-year clinical outcomes, including recurrent angina. (NeoVas Bioresorbable Coronary Scaffold Randomized Controlled Trial; NCT02305485)