CBS 2019
CBSMD教育中心
中 文

光学相关断层扫描

Abstract

Recommended Article

Device specificity of vascular healing following implantation of bioresorbable vascular scaffolds and bioabsorbable polymer metallic drug-eluting stents in human coronary arteries: the ESTROFA OCT BVS vs. BP-DES study Volumetric characterization of human coronary calcification by frequency-domain optical coherence tomography The Relation Between Optical Coherence Tomography-Detected Layered Pattern and Acute Side Branch Occlusion After Provisional Stenting of Coronary Bifurcation Lesions Clinical Predictors for Lack of Favorable Vascular Response to Statin Therapy in Patients With Coronary Artery Disease: A Serial Optical Coherence Tomography Study Optical coherence tomography versus intravascular ultrasound to evaluate coronary artery disease and percutaneous coronary intervention Lipid-rich plaque and myocardial perfusion after successful stenting in patients with non-ST-segment elevation acute coronary syndrome: an optical coherence tomography study Coronary Atherosclerosis T1-Weighed Characterization With Integrated Anatomical Reference: Comparison With High-Risk Plaque Features Detected by Invasive Coronary Imaging Optical Coherence Tomography-Guided Percutaneous Coronary Intervention in ST-Segment-Elevation Myocardial Infarction: A Prospective Propensity-Matched Cohort of the Thrombectomy Versus Percutaneous Coronary Intervention Alone Trial

Clinical TrialVolume 11, Issue 3, February 2018

JOURNAL:JACC Cardiovasc Interv. Article Link

A Randomized Trial Comparing the NeoVas Sirolimus-Eluting Bioresorbable Scaffold and Metallic Everolimus-Eluting Stents

YL Han, B Xu, NeoVas Randomized Controlled Trial Investigators Keywords: bioresorbable scaffolds; drug-eluting stent(s); randomized controlled trial

ABSTRACT



OBJECTIVES - The authors sought to evaluate the safety and effectiveness of the NeoVas bioresorbable scaffold (BRS) compared with metallic drug-eluting stents.



BACKGROUND -  BRS have the potential to improve very late outcomes compared with metallic drug-eluting stents, but some BRS have been associated with increased rates of device thrombosis before complete bioresorption. NeoVas is a new poly-L-lactic acid BRS that elutes sirolimus from a poly-D, L-lactide coating.



METHODS -  Eligible patients with a single de novo native coronary artery lesion with a reference vessel diameter 2.5 to 3.75 mm and a lesion length ≤20 mm were randomized 1:1 to NeoVas BRS versus cobalt-chromium everolimus-eluting stents (CoCr-EES). Angiographic follow-up was performed in all patients at 1 year. The primary endpoint was angiographic in-segment late loss (LL), and the major secondary endpoint was the rate of angina. Baseline and follow-up optical coherence tomography and fractional flow reserve were performed in a pre-specified subgroup of patients.



RESULTS -  The authors randomized 560 patients at 32 centers to treatment with NeoVas (n = 278) versus CoCr-EES (n = 282). One-year in-segment LL with NeoVas and CoCr-EES were 0.14 ± 0.36 mm versus 0.11 ± 0.34 mm (difference 0.03 mm; upper 1-sided 97.5% confidence interval 0.09 mm; pnoninferiority < 0.0001; psuperiority = 0.36). Clinical outcomes at 1 year were similar in the 2 groups, as were the rates of recurrent angina (27.9% vs. 32.1%; p = 0.26). Optical coherence tomography at 1 year demonstrated a higher proportion of covered struts (98.7% vs. 96.2%; p < 0.001), less strut malapposition (0% vs. 0.6%; p <0.001), and a smaller minimal lumen area (4.71 ± 1.64 vs. 6.00 ± 2.15 mm2; p < 0.001) with NeoVas compared with CoCr-EES respectively, with nonsignificant differences in fractional flow reserve (0.89 ± 0.08 vs. 0.91 ± 0.06; p = 0.07).



CONCLUSIONS - The NeoVas BRS was noninferior to CoCr-EES for the primary endpoint of 1-year angiographic in-segment LL, and resulted in comparable 1-year clinical outcomes, including recurrent angina. (NeoVas Bioresorbable Coronary Scaffold Randomized Controlled Trial; NCT02305485)