Clinical use of intracoronary imaging. Part 1: guidance and optimization of coronary interventions. An expert consensus document of the European Association of Percutaneous Cardiovascular Interventions: Endorsed by the Chinese Society of Cardiology A new optical coherence tomography-based calcium scoring system to predict stent underexpansion Neoatherosclerosis in Patients With Coronary Stent Thrombosis: Findings From Optical Coherence Tomography Imaging (A Report of the PRESTIGE Consortium) Characteristics of stent thrombosis in bifurcation lesions analysed by optical coherence tomography Angiography Alone Versus Angiography Plus Optical Coherence Tomography to Guide Percutaneous Coronary Intervention: Outcomes From the Pan-London PCI Cohort Impact of low tissue backscattering by optical coherence tomography on endothelial function after drug-eluting stent implantation Myocardial Blood Flow and Coronary Flow Reserve During 3 Years Following Bioresorbable Vascular Scaffold Versus Metallic Drug-Eluting Stent Implantation: The VANISH Trial Coronary Artery Intraplaque Microvessels by Optical Coherence Tomography Correlate With Vulnerable Plaque and Predict Clinical Outcomes in Patients With Ischemic Angina

Clinical TrialVolume 11, Issue 3, February 2018

JOURNAL:JACC Cardiovasc Interv. Article Link

A Randomized Trial Comparing the NeoVas Sirolimus-Eluting Bioresorbable Scaffold and Metallic Everolimus-Eluting Stents

YL Han, B Xu, NeoVas Randomized Controlled Trial Investigators Keywords: bioresorbable scaffolds; drug-eluting stent(s); randomized controlled trial

ABSTRACT



OBJECTIVES - The authors sought to evaluate the safety and effectiveness of the NeoVas bioresorbable scaffold (BRS) compared with metallic drug-eluting stents.



BACKGROUND -  BRS have the potential to improve very late outcomes compared with metallic drug-eluting stents, but some BRS have been associated with increased rates of device thrombosis before complete bioresorption. NeoVas is a new poly-L-lactic acid BRS that elutes sirolimus from a poly-D, L-lactide coating.



METHODS -  Eligible patients with a single de novo native coronary artery lesion with a reference vessel diameter 2.5 to 3.75 mm and a lesion length ≤20 mm were randomized 1:1 to NeoVas BRS versus cobalt-chromium everolimus-eluting stents (CoCr-EES). Angiographic follow-up was performed in all patients at 1 year. The primary endpoint was angiographic in-segment late loss (LL), and the major secondary endpoint was the rate of angina. Baseline and follow-up optical coherence tomography and fractional flow reserve were performed in a pre-specified subgroup of patients.



RESULTS -  The authors randomized 560 patients at 32 centers to treatment with NeoVas (n = 278) versus CoCr-EES (n = 282). One-year in-segment LL with NeoVas and CoCr-EES were 0.14 ± 0.36 mm versus 0.11 ± 0.34 mm (difference 0.03 mm; upper 1-sided 97.5% confidence interval 0.09 mm; pnoninferiority < 0.0001; psuperiority = 0.36). Clinical outcomes at 1 year were similar in the 2 groups, as were the rates of recurrent angina (27.9% vs. 32.1%; p = 0.26). Optical coherence tomography at 1 year demonstrated a higher proportion of covered struts (98.7% vs. 96.2%; p < 0.001), less strut malapposition (0% vs. 0.6%; p <0.001), and a smaller minimal lumen area (4.71 ± 1.64 vs. 6.00 ± 2.15 mm2; p < 0.001) with NeoVas compared with CoCr-EES respectively, with nonsignificant differences in fractional flow reserve (0.89 ± 0.08 vs. 0.91 ± 0.06; p = 0.07).