荐读文献

Spontaneous Coronary Artery Dissection: JACC State-of-the-Art Review Management of Percutaneous Coronary Intervention Complications: Algorithms From the 2018 and 2019 Seattle Percutaneous Coronary Intervention Complications Conference Generalizing Intensive Blood Pressure Treatment to Adults With Diabetes Mellitus The Future of Cardiovascular Computed Tomography Advanced Analytics and Clinical Insights Randomized Comparison of Ridaforolimus-Eluting and Zotarolimus-Eluting Coronary Stents 2-Year Clinical Outcomes: From the BIONICS and NIREUS Trials Management of Patients With NSTE-ACS: A Comparison of the Recent AHA/ACC and ESC Guidelines Syncope After Percutaneous Coronary Intervention Position paper of the EACVI and EANM on artificial intelligence applications in multimodality cardiovascular imaging using SPECT/CT, PET/CT, and cardiac CT Coronary Artery Calcium Is Associated with Left Ventricular Diastolic Function Independent of Myocardial Ischemia Influence of LDL-Cholesterol Lowering on Cardiovascular Outcomes in Patients With Diabetes Mellitus Undergoing Coronary Revascularization

Original Research14 September 2021

JOURNAL:Pacing Clin Electrophysiol. Article Link

Significantly less inappropriate shocks in ischemic patients compared to non-ischemic patients: The S-ICD experience of a high volume single-center

E Oosterwerff, A Adiyaman, A Elvan et al. Keywords: S-ICD; ischemic cardiomyopathy; non-ischemic cardiomyopathy; inappropriate shocks

ABSTRACT

BACKGROUND - The subcutaneous cardioverter-defibrillator (S-ICD) continues to be preferentially used in relatively young patients, with less advanced heart disease.

 

OBJECTIVE - We, therefore, studied the short and long-term efficacy and safety of the S-ICD in subgroups of patients, which are underreported at present.

 

METHODS - A total of 218 patients between November 2010 and February 2019 undergoing S-ICD with a follow up of at least 6 months implantation were included in a prospective registry. Mean follow up was 38 months.

 

RESULTS - The most common indication for S-ICD implantation was ischemic cardiomyopathy (n = 106, 49%). Complication rate needing invasive intervention was 9% (n = 21). Appropriate shock rate in patients with an S-ICD was 3.5%/year. A total of 30 inappropriate shocks (IAS) occurred in 19 patients (8.7%; 2.7%/year). The proportion of appropriate and inappropriate shock rates in patients with different cardiomyopathies shows remarkable variances. There were significant more IAS (3.6%/year vs. 1.7%/year,p = .048) in patients with non-ischemic cardiomyopathy versus patients with ischemic cardiomyopathy. Multivariate analysis identified, besides type of cardiomyopathy, atrial fibrillation (AF) as predictor for IAS.

 

CONCLUSION - In this real-world prospective registry we analyzed S-ICD performance in the more traditional ICD patient. Patients with ischemic cardiomyopathy had significantly less inappropriate therapy compared to patients with non-ischemic cardiomyopathy and appear to be appropriate patients for this type of device.