CBS 2019
CBSMD教育中心
English

光学相关断层扫描

科研文章

荐读文献

Clinical Predictors for Lack of Favorable Vascular Response to Statin Therapy in Patients With Coronary Artery Disease: A Serial Optical Coherence Tomography Study Covering our tracks – optical coherence tomography to assess vascular healing Optical coherence tomography versus intravascular ultrasound to evaluate coronary artery disease and percutaneous coronary intervention Assessment of the coronary calcification by optical coherence tomography A Survey on Coronary Atherosclerotic Plaque Tissue Characterization in Intravascular Optical Coherence Tomography Intravascular optical coherence tomography Volumetric characterization of human coronary calcification by frequency-domain optical coherence tomography Characteristics of abnormal post-stent optical coherence tomography findings in hemodialysis patients Optical coherence tomography and C-reactive protein in risk stratification of acute coronary syndromes Consensus standards for acquisition, measurement, and reporting of intravascular optical coherence tomography studies: a report from the International Working Group for Intravascular Optical Coherence Tomography Standardization and Validation

Clinical Trial2013 Mar;6(3):228-36.

JOURNAL:JACC Cardiovasc Interv. Article Link

Optical coherence tomography versus intravascular ultrasound to evaluate coronary artery disease and percutaneous coronary intervention

Bezerra HG, Attizzani GF, Sirbu V et al. Keywords: IVUS; OCT; PCI

ABSTRACT


OBJECTIVESWe compared intravascular ultrasound (IVUS) and 2 different generations of optical coherence tomography (OCT)-time-domain OCT (TD-OCT) and frequency-domain OCT (FD-OCT)-for the assessment of coronary disease and percutaneous coronary intervention (PCI) using stents.


BACKGROUND - OCT is a promising light-based intravascular imaging modality with higher resolution than IVUS. However, the paucity of data on OCT image quantification has limited its application in clinical practice.

METHODS - A total of 227 matched OCT and IVUS pull backs were studied. One hundred FD-OCT and IVUS pull backs in nonstented (n = 56) and stented (n = 44) vessels were compared. Additionally, 127 matched TD-OCT and IVUS images were compared in stented vessels.

RESULTS - FD-OCT depicted more severe native coronary disease than IVUS; minimal lumen area (MLA) was 2.33 ± 1.56 mm(2) versus 3.32 ± 1.92 mm(2), respectively (p < 0.001). Reference vessel dimensions were equivalent between FD-OCT and IVUS in both native and stented coronaries, but TD-OCT detected smaller reference lumen size compared with IVUS. Immediately post-PCI, in-stent MLAs were similar between FD-OCT and IVUS, but at follow-up, both FD-OCT and TD-OCT detected smaller MLAs than did IVUS, likely due to better detection of neointimal hyperplasia (NIH). Post-PCI malapposition and tissue prolapse were more frequently identified by FD-OCT.

CONCLUSIONS - FD-OCT generates similar reference lumen dimensions but higher degrees of disease severity and NIH, as well as better detection of malapposition and tissue prolapse compared with IVUS. First-generation TD-OCT was associated with smaller reference vessel dimensions compared with IVUS.

Copyright © 2013 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.