CBS 2019
CBSMD教育中心
中 文

IVUS Guidance

Abstract

Recommended Article

The impact of intravascular ultrasound guidance during drug eluting stent implantation on angiographic outcomes IVUS in bifurcation stenting: what have we learned? Intravascular Ultrasound Guidance vs. Angiographic Guidance in Primary Percutaneous Coronary Intervention for ST-Segment Elevation Myocardial Infarction - Long-Term Clinical Outcomes From the CREDO-Kyoto AMI Registry Usefulness of minimum stent cross sectional area as a predictor of angiographic restenosis after primary percutaneous coronary intervention in acute myocardial infarction (from the HORIZONS-AMI Trial IVUS substudy) Intravascular ultrasound findings of early stent thrombosis after primary percutaneous intervention in acute myocardial infarction: a Harmonizing Outcomes with Revascularization and Stents in Acute Myocardial Infarction (HORIZONS-AMI) substudy Intravascular Ultrasound-Derived Virtual Fractional Flow Reserve for the Assessment of Myocardial Ischemia Usefulness of intravascular ultrasound to predict outcomes in short-length lesions treated with drug-eluting stents Clinical impact of PCSK9 inhibitor on stabilization and regression of lipid-rich coronary plaques: a near-infrared spectroscopy study

Original Research 2018 Sep 30. [Epub ahead of print]

JOURNAL:Catheter Cardiovasc Interv. Article Link

Novel predictor of target vessel revascularization after coronary stent implantation: Intraluminal intensity of blood speckle on intravascular ultrasound

Saito Y, Kitahara H, Okuya Y et al. Keywords: intravascular ultrasound; percutaneous coronary intervention; target vessel revascularization

ABSTRACT


BACKGROUND - The difference in intraluminal intensity of blood speckle (IBS) on integrated backscatter-intravascular ultrasound (IB-IVUS) across the coronary artery stenosis (i.e., ΔIBS) has been reported to negatively correlate with fractional flow reserve. Fractional flow reserve after coronary stenting is known as a predictor of target vessel revascularization (TVR). However, the relation between ΔIBS and TVR is unclear.


METHODS - Seven hundred and three vessels which underwent percutaneous coronary intervention with stents were screened. Vessels without IVUS-guidance and follow-up information were excluded. Intraluminal IBS values were measured using IB-IVUS in cross-sections at the ostium of the target vessel and at the distal reference of implanted stent. ΔIBS was calculated as (distal IBS) - (ostium IBS).


RESULTS - A total of 393 vessels were included. Mean ΔIBS at postprocedure was 6.22 ± 5.65. During the follow-up period (11.2 ± 3.1 months), 24 cases (6.1%) had TVR. ΔIBS was significantly greater in the vessels with TVR than in those without (11.10 ± 5.93 vs. 5.90 ± 5.49, P <0.001). In receiver operating characteristic curve analysis, ΔIBS significantly predicted TVR (AUC 0.74, best cut-off value 8.24, P < 0.001). Multiple logistic regression analysis showed use of drug eluting stent and ΔIBS ≥ 8.24 as independent predictors of TVR.


CONCLUSIONS - ΔIBS at postprocedure was significantly associated with TVR. IVUS may be able to predict TVR by physiological assessment with measurement of ΔIBS.


© 2018 Wiley Periodicals, Inc.