Impact of plaque components on no-reflow phenomenon after stent deployment in patients with acute coronary syndrome: a virtual histology-intravascular ultrasound analysis In vivo intravascular ultrasound-derived thin-cap fibroatheroma detection using ultrasound radiofrequency data analysis Differential prognostic effect of intravascular ultrasound use according to implanted stent length Comprehensive intravascular ultrasound assessment of stent area and its impact on restenosis and adverse cardiac events in 403 patients with unprotected left main disease Histopathologic validation of the intravascular ultrasound diagnosis of calcified coronary artery nodules Is intravascular ultrasound beneficial for percutaneous coronary intervention of bifurcation lesions? Evidence from a 4,314-patient registry Intravascular ultrasound assessment of the effects of rotational atherectomy in calcified coronary artery lesions Intravascular Ultrasound Guidance Reduces Cardiac Death and Coronary Revascularization in Patients Undergoing Drug-Eluting Stent Implantation: Results From a Meta-Analysis of 9 Randomized Trials and 4724 Patients

Original Research 2018 Sep 30. [Epub ahead of print]

JOURNAL:Catheter Cardiovasc Interv. Article Link

Novel predictor of target vessel revascularization after coronary stent implantation: Intraluminal intensity of blood speckle on intravascular ultrasound

Saito Y, Kitahara H, Okuya Y et al. Keywords: intravascular ultrasound; percutaneous coronary intervention; target vessel revascularization

ABSTRACT


BACKGROUND - The difference in intraluminal intensity of blood speckle (IBS) on integrated backscatter-intravascular ultrasound (IB-IVUS) across the coronary artery stenosis (i.e., ΔIBS) has been reported to negatively correlate with fractional flow reserve. Fractional flow reserve after coronary stenting is known as a predictor of target vessel revascularization (TVR). However, the relation between ΔIBS and TVR is unclear.


METHODS - Seven hundred and three vessels which underwent percutaneous coronary intervention with stents were screened. Vessels without IVUS-guidance and follow-up information were excluded. Intraluminal IBS values were measured using IB-IVUS in cross-sections at the ostium of the target vessel and at the distal reference of implanted stent. ΔIBS was calculated as (distal IBS) - (ostium IBS).


RESULTS - A total of 393 vessels were included. Mean ΔIBS at postprocedure was 6.22 ± 5.65. During the follow-up period (11.2 ± 3.1 months), 24 cases (6.1%) had TVR. ΔIBS was significantly greater in the vessels with TVR than in those without (11.10 ± 5.93 vs. 5.90 ± 5.49, P <0.001). In receiver operating characteristic curve analysis, ΔIBS significantly predicted TVR (AUC 0.74, best cut-off value 8.24, P < 0.001). Multiple logistic regression analysis showed use of drug eluting stent and ΔIBS ≥ 8.24 as independent predictors of TVR.


CONCLUSIONS - ΔIBS at postprocedure was significantly associated with TVR. IVUS may be able to predict TVR by physiological assessment with measurement of ΔIBS.