Physiologic Characteristics and Clinical Outcomes of Patients With Discordance Between FFR and iFR Sex Differences in Instantaneous Wave-Free Ratio or Fractional Flow Reserve–Guided Revascularization Strategy Real-world clinical utility and impact on clinical decision-making of coronary computed tomography angiography-derived fractional flow reserve: lessons from the ADVANCE Registry Diagnostic accuracy of fractional flow reserve from anatomic CT angiography Comparison of Coronary Computed Tomography Angiography, Fractional Flow Reserve, and Perfusion Imaging for Ischemia Diagnosis Diagnostic performance of noninvasive fractional flow reserve derived from coronary computed tomography angiography in suspected coronary artery disease: the NXT trial (Analysis of Coronary Blood Flow Using CT Angiography: Next Steps) Experimental basis of determining maximum coronary, myocardial, and collateral blood flow by pressure measurements for assessing functional stenosis severity before and after percutaneous transluminal coronary angioplasty Fractional Flow Reserve-Guided Complete Revascularization Improves the Prognosis in Patients With ST-Segment-Elevation Myocardial Infarction and Severe Nonculprit Disease: A DANAMI 3-PRIMULTI Substudy (Primary PCI in Patients With ST-Elevation Myocardial Infarction and Multivessel Disease: Treatment of Culprit Lesion Only or Complete Revascularization)

Original ResearchVolume 11, Issue 15, August 2018

JOURNAL:JACC Cardiovasc Interv. Article Link

New Volumetric Analysis Method for Stent Expansion and its Correlation With Final Fractional Flow Reserve and Clinical Outcome An ILUMIEN I Substudy

D Nakamura, W Wijns, MJ Price et al. Keywords: fractional flow reserve; optical coherence tomography; restenosis; stent expansion

ABSTRACT


OBJECTIVES - This study sought to compare conventional methodology (CM) with a newly described optical coherence tomography (OCT)-derived volumetric stent expansion analysis in terms of fractional flow reserve (FFR)-derived physiology and device-oriented composite endpoints (DoCE).


BACKGROUND - The analysis of coronary stent expansion with intracoronary imaging has used CM that relies on the analysis of selected single cross-sections for several decades. The introduction of OCT with its ability to perform semiautomated volumetric analysis opens opportunities to redefine optimal stent expansion.

METHODS - A total of 291 lesions treated with post-stent OCT and FFR were enrolled. The expansion index was calculated by using a novel volumetric algorithm and was defined as: ([actual lumen area / ideal lumen area] × 100) for each frame of the stented segment. The minimum expansion index (MEI) was defined as the minimum value of expansion index along the entire stented segment. MEI and conventional lumen expansion metrics were compared for the ability to predict post-stent low FFR (<0.90) and DoCE at 1 year.

RESULTS - There was a stronger correlation between MEI and final FFR, compared with CM and final FFR (r = 0.690; p < 0.001) versus (r = 0.165; p = 0.044). MEI was significantly lower in patients with DoCE than those without DoCE (72.18 ± 8.23% vs. 81.48 ± 11.03%; p < 0.001), although stent expansion by CM was similar between patients with and without DoCE (85.05 ± 22.19% and 83.73 ± 17.52%; p = 0.858), respectively.

CONCLUSIONS - OCT analysis of stent expansion with a newly described volumetric method, but not with CM, yielded data that were predictive of both an acute improvement in FFR-derived physiology and DoCE.