Can We Use the Intrinsic Left Ventricular Delay (QLV) to Optimize the Pacing Configuration for Cardiac Resynchronization Therapy With a Quadripolar Left Ventricular Lead? Frequency, predictors, and prognosis of ejection fraction improvement in heart failure: an echocardiogram-based registry study Diagnostic performance of congestion score index evaluated from chest radiography for acute heart failure in the emergency department: A retrospective analysis from the PARADISE cohort 3D Printing and Heart Failure: The Present and the Future Percutaneous Coronary Intervention vs Coronary Artery Bypass Grafting in Patients With Left Main Coronary Artery Stenosis A Systematic Review and Meta-analysis 2016 ACC/AHA guideline focused update on duration of dual antiplatelet therapy in patients with coronary artery disease: A report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines Improving the Use of Primary Prevention Implantable Cardioverter-Defibrillators Therapy With Validated Patient-Centric Risk Estimates Association of White Matter Hyperintensities and Cardiovascular Disease: The Importance of Microcirculatory Disease

Clinical Trial2018 Mar;11(3):e005912.

JOURNAL:Circ Arrhythm Electrophysiol. Article Link

Can We Use the Intrinsic Left Ventricular Delay (QLV) to Optimize the Pacing Configuration for Cardiac Resynchronization Therapy With a Quadripolar Left Ventricular Lead?

van Everdingen WM, Zweerink A, Cramer MJ et al. Keywords: cardiac resynchronization therapy; heart failure; hemodynamics; humans; stroke volume

ABSTRACT


BACKGROUNDPrevious studies indicated the importance of the intrinsic left ventricular (LV) electric delay (QLV) for optimal benefit to cardiac resynchronization therapy. We investigated the use of QLV for achieving optimal acute hemodynamic response to cardiac resynchronization therapy with a quadripolar LV lead.


METHODS AND RESULTS - Forty-eight heart failure patients with a left bundle branch block were prospectively enrolled (31 men; age, 66±10 years; LV ejection fraction, 28±8%; QRS duration, 176±14 ms). Immediately after cardiac resynchronization therapy implantation, invasive LV pressure-volume loops were recorded during biventricular pacing with each separate electrode at 4 atrioventricular delays. Acute cardiac resynchronization therapy response, measured as change in stroke work (Δ%SW) compared with intrinsic conduction, was related to intrinsic interval between Q on the ECG and LV sensing delay (QLV), normalized for QRS duration (QLV/QRSd), and electrode position. QLV/QRSd was 84±9% and variation between the 4 electrodes 9±5%. Δ%SW was 89±64% and varied by 39±36% between the electrodes. In univariate analysis, an anterolateral or lateral electrode position and a high QLV/QRSd had a significant association with a large Δ%SW (all P <0.01). In a combined model, only QLV/QRSd remained significantly associated with Δ%SW (P<0.05). However, a direct relation between QLV/QRSd and Δ%SW was only seen in 24 patients, whereas 24 patients showed an inverse relation.

CONCLUSIONS - The large variation in acute hemodynamic response indicates that the choice of the stimulated electrode on a quadripolar lead is important. Although QLV/QRSd was associated with acute hemodynamic response at group level, it cannot be used to select the optimal electrode in the individual patient.