Does calcium burden impact culprit lesion morphology and clinical results? An ADAPT-DES IVUS substudy Intraluminal Intensity of Blood Speckle on Intravascular Ultrasound, a Novel Predictor of Periprocedural Myocardial Injury After Coronary Stenting Subclinical Atherosclerosis Burden by 3D Ultrasound in Mid-Life: The PESA Study Intravascular ultrasound-guided drug-eluting stent implantation is associated with improved clinical outcomes in patients with unstable angina and complex coronary artery true bifurcation lesions The outcomes of intravascular ultrasound-guided drug-eluting stent implantation among patients with complex coronary lesions: a comprehensive meta-analysis of 15 clinical trials and 8,084 patients Intravascular Ultrasound Guidance Reduces Cardiac Death and Coronary Revascularization in Patients Undergoing Drug-Eluting Stent Implantation: Results From a Meta-Analysis of 9 Randomized Trials and 4724 Patients IVUS Guidance for Coronary Revascularization: When to Start, When to Stop? Incidence and Clinical Outcomes of Stent Fractures on the Basis of 6,555 Patients and 16,482 Drug-Eluting Stents From 4 Centers

Original Research25 May 2021

JOURNAL:Eurointervention. Article Link

Incidence and Standardized Definitions of Mitral Valve Leaflet Adverse Events After Transcatheter Mitral Valve Repair: the EXPAND Study

FM Asch, SH Little, GB Mackensen et al. Keywords: mitral valve leaflet adverse events; LAE; definition; MitraClip;

ABSTRACT

BACKGROUND - An independent panel of experts reviewed all investigator-reported cases of mitral valve leaflet adverse events (LAE) after MitraClipTM NTR/XTR in the EXPAND Study.


AIMS - We aimed to report the findings of the expert panel and standardize definitions for LAE.


METHODS - Standard definitions for different types of LAE were formulated and events adjudicated after detailed review by the expert panel.


RESULTS - Enrolling centers reported LAE in 35 cases, 11 leaflet injuries (9 tear, 2 perforation) and 24 single leaflet device attachment (SLDA). The panel confirmed LAE in 20 cases (2.0% incidence), 18 patients had SLDA and 4 had leaflet injury (2 cases had both SLDA and injury). Leaflet injury occurred during device implant and resulted in surgical valve replacement or death. SLDA-alone events were identified during implant (n=2), pre-discharge (7) or at 30 days of follow-up (7) and were resolved (£ 2+ residual MR) with additional clips in 75% of cases.


CONCLUSIONS - Mitral valve repair with MitraClipTM NTR/XTR is safe. The rate of LAE is lower than previously reported using older generation devices. The proposed definitions and findings will help differentiate leaflet injury from inadequate leaflet insertion and SLDA, and provide guidance to consistently diagnose LAE post MitraClipTM.